Measuring range: 0.5 – 20.0 mg/l PO
1.5 – 61.3 mg/l PO
1.1 – 45.8 mg/l P
Check the pH of the
Select method 3 8 1 .
sample, specified range:
pH 0 – 10.
If required, add dilute
sulfuric acid drop by
drop to adjust the pH.
Add 1 dose of P-3K
Shake the cell vigor-
using the blue dose-
ously to dissolve the
metering cap, close with
solid substance.
the screw cap.
Release 08/2014 - Spectroquant
Phosphate
Determination of orthophosphate
-P
16-mm cell
4
16-mm cell
4
O
16-mm cell
2
5
Fill approx. 10 ml of
distilled water into an
empty 16-mm cell (do
not add any reagents!),
close with the screw
cap.
(Blank cell)
Reaction time:
5 minutes
Press Enter to start
the countdown.
®
Multy
Pipette 1.0 ml of the
sample into a reaction
cell, close with the screw
cap, and mix.
Insert the blank cell into
the cell compartment.
Press Zero .
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) we recommended to use
®
Spectroquant
CombiCheck 20 and 80, Cat.No. 114675
and 114738.
Ready-for-use phosphate standard solution Certipur
Cat.No. 119898, concentration 1000 mg/l PO
be used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck) is highly recommended.
114729
Cell Test
Add 5 drops of P-2K,
close with the screw
cap, and mix.
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
®
,
3–
, can also
4