Millipore Spectroquant Multy Operating Manual page 197

Colorimeter
Table of Contents

Advertisement

Measuring range: 50 – 3000 mg/l volatile organic acid
71 – 4401 mg/l volatile organic acid
Check the pH of the
Select method 5 3 1 .
sample, specified range:
pH 2 – 12.
Heat both cells in the
Add to each cell 1.0 ml of
thermoreactor at 100 °C
OA-2K with pipette.
for 15 minutes. Then cool
to room temperature
under running water.
Insert the blank cell into
Insert the cell containing
the cell compartment.
the sample into the cell
Align the mark on the
compartment. Align the
cell with that on the pho-
mark on the cell with
tometer. Press Zero .
that on the photometer.
Press Test .
Release 08/2014 - Spectroquant
Volatile Organic Acids
Pipette 0.50 ml each of
OA-1K into two reaction
cells.
Add to each cell 1.0 ml
of OA-3K with pipette,
close with the screw cap,
and mix.
®
Multy
(calculated as acetic acid)
(calculated as butyric acid)
Add to one cell 0.50 ml
of the sample with pi-
pette, close with the
screw cap, and mix.
Add to each cell 1.0 ml
of OA-4K with pipette,
close with the screw cap,
and mix.
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) a standard solution must
be prepared from sodium acetate anhydrous, Cat.No.
106268 (see section "Standard solutions").
101749
Cell Test
16-mm cell
16-mm cell
Add to the second cell
0.50 ml of distilled water
with pipette, close with
the screw cap, and mix.
(Blank cell)
Reaction time:
1 minute
Press Enter to start
the countdown.