Millipore Spectroquant Multy Operating Manual page 251

Colorimeter
Table of Contents

Advertisement

Go to "Options > Communication parameters" and set the
baud rate to 19200 and ComPort to "AUTO" (or the number of
the connected COM port, e.g. Com-Port 1). Then click on "OK".
Fig. 2
After this go to the menu item "File > Open..". for HexLoad to
load the software update (*.hex file). Now switch the colori-
meter on. When a connection to HexLoad has been established,
the display of the Spectroquant® Multy Colorimeters remains
blank.
HexLoad should now look similar to the example below, although
the figures actually shown may vary:
Fig. 3
It is essential that:
under "File" in the top lefthand corner the message
"No data loaded" has changed and been replaced by
values similar to those shown above; and
under "Target" in the top righthand corner values
(in blue type) are now shown.
Chapter 5.5 | Page 3 of 5