5.6.5 Initializing the user-method system
(concentration and polynomial)
A loss of power results in incoherent data for stored user
methods. The user-method system must then by initialized with
this mode function to return it to a default standard.
Press the keys [Mode] [6] [9] in succession.
Confirm your selection by pressing [Enter].
The display now shows:
Confirm your selection by pressing [Enter].
The display now shows the prompt message:
• Press key [1] to start the initializing procedure.
m
• Press key [0] to abort the initializing procedure.
The colorimeter automatically returns to the mode menu.
Attention:
All stored concentration methods and polynomials are
deleted by the initializing procedure!
<User m. init.>
Start:
initialising?
YES: 1
NO: 0
Chapter 5.6 | Page 13 of 13