Measuring range: 0.05– 6.00 mg/l Mn
Attention: In contrast to the instructions given in the package insert the sample volume as well as the reagent volume
must be doubled.
Check the pH of the
Select method 2 8 2 .
sample, specified range:
pH 2 – 7.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Add 4 drops of Mn-2,
Reaction time:
close with the screw
2 minutes
cap, and mix.
Press Enter to start
Check the pH, specified
the countdown.
pH: approx. 11.5.
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Release 08/2014 - Spectroquant
Manganese
24-mm cell
Fill approx. 10 ml of
distilled water into a
24-mm cell (do not add
any reagents!), close
with the screw cap.
(Blank cell)
Add 4 drops of Mn-3,
close with the screw
cap, and mix.
®
Multy
Pipette 10 ml of the
sample into a 24-mm
cell.
Reaction time:
2 minutes
Press Enter to start
the countdown.
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) we recommended to use
®
Spectroquant
CombiCheck 30, Cat.No. 114677.
Use 10 ml R-1 instead of the sample.
Ready-for-use manganese standard solution Certipur
Cat.No. 119789, concentration 1000 mg/l Mn, can also be
used after diluting accordingly.
To check for sample-dependent effects the use of addition
solutions (e.g. in CombiCheck 30) is highly recommended.
Use 10 ml sample + 0.1 ml R-2.
114770
Test
Add 8 drops of Mn-1,
close with the screw
cap, and mix.
Check the pH, specified
pH: approx. 11.5.
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .
®
,