Millipore Spectroquant Multy Operating Manual page 154

Colorimeter
Table of Contents

Advertisement

Measuring range: pH 6.4 – 8.8
Select method 3 6 0 .
Fill approx. 10 ml of
distilled water into an
empty 16-mm cell (do
not add any reagents!),
close with the screw
cap.
(Blank cell)
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Release 08/2014 - Spectroquant
pH
16-mm cell
Pipette 10 ml of the
sample into a round
cell.
®
Multy
Add 4 drops of pH-1,
close with the screw
cap, and mix. Attention!
The reagent bottle must
be held vertically by all
means!
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) buffer solution pH 7.00
®
Certipur
, Cat.No. 109407, can be used.
101744
Cell Test
Insert the blank cell into
the cell compartment.
Press Zero .