Trouble-Shooting; User Messages On The Display / Error Messages - Millipore Spectroquant Multy Operating Manual

Colorimeter
Table of Contents

Advertisement

4.1
User messages on the display
/ Error messages
Display message: ADU Err 01: xx
Measure: Initialize LED parameters
Please carry out the following steps to set and save the LED
brightness parameters:
Test the software version of the unit by pressing the keys
[Mode] [9] [1].
The version should be V012.00x.3.003.zzz, with zzz = 003 or
higher. If you have a lower software version you should per-
form a software update (see chapter 5.5, "Software update via
Internet"). To do this download the current version for the
Spectroquant® Multy from our website
(www.service-test-kits.com).
The cell compartment must be clean and dry.
Fill one cell with distilled water and place it in the cell com-
partment. It is irrelevant whether you use a 16-mm or 24-mm
round cell.
Take care that the unit is NOT exposed to direct sunlight or a
bright spotlight.
Keep the [Mode] key pressed, switch off the unit
and switch it back on while pressing the [Mode] key.

Trouble-shooting

(xx = 01, 02, 10, or 20)
en
39

Advertisement

Table of Contents
loading

Table of Contents