The five-digit item number (e.g. 14541) gives the five digits
in the middle of the catalogue/ordering number 1.XXXXX.0001,
in this case 1.14541.0001. In some cases in which the assign-
ment is self-evident (e.g. monochloramine)
or else in which all Spectroquant® tests available for one
parameter can be used (e.g. chlorine), this number is not shown.
For an overview of all programmed methods please refer to the
included CD, section 5.1, "Overview of preprogrammed methods
and analytical procedures".
2.2

Measuring with test kits

A detailed description of the procedure for the selected method
is given on the provided CD in section 5.1, "Overview of prepro-
grammed methods and analytical procedures". The procedures
may differ slightly from those described in the respective pack-
age inserts.
After selecting the method, prepare the blank and sample for
measurement.
In the case of analysis specifications in which reaction times
must be observed, a timer (countdown) is integrated in
the method programme. (In such cases the cells must not be
inserted into the measurement compartment.)
After the method has been selected the display shows:
Example: Method 90, Bromine 00605
If you wish to exit the menu at this stage, simply press the
[Enter] key twice (= abort countdown) and then the [Esc]
key once.
After the method has been selected, the countdown function
is started by pressing the [Enter] key. The remaining time is
shown continuously. An acoustic signal is emitted in the last
10 seconds before the countdown expires. After the countdown
has run out, proceed as described in the analysis procedures.
In some methods there are several reaction times that have
to be considered; these are shown and processed in the proper
sequence.
90 Bromine 00605
0.10-5.00 mg/l Br2
Count-down 1
1:00
Start:
en