Millipore Spectroquant Multy Operating Manual page 257

Colorimeter
Table of Contents

Advertisement

Measurement mode with standards of known
concentrations
The display now shows:
Prepare zero and press [Zero].
Note
Use distilled water or a reagent blank.
The display now shows:
Enter the concentration of the first standard;
e.g.: [0] [,] [0] [5] for 0.05
.
• Back with the key [Esc].
• Backout the entry with the key [F1].
Confirm your selection by pressing [Enter].
The display now shows:
Prepare the first standard and press [Test].
The display shows the entered value and the measured
absorbance value:
Confirm your selection by pressing [Enter].
<User concentr.>
Prepare Zero
Press ZERO
<User concentr.>
Zero accepted
S1: +_____
ESC
F1
<User concentr.>
S1: 0.05 mg/l
Prepare
Press TEST
S1: 0.05 mg/l
mAbs: 12
Chapter 5.6 | Page 4 of 13