Millipore Spectroquant Multy Operating Manual page 263

Colorimeter
Table of Contents

Advertisement

The display now shows:
Press
the number keys to select the desired resolution, e.g.: [2]
for 0.1.
Note
Please adjust the desired resolution according to the fol-
lowing criteria:
Range
max. Resolution
0.000 ...9,999
0.001
10.00 ...99,99
0.01
100.0 ...999,9
0.1
1000 ...9999
1
The display now shows:
The colorimeter automatically returns to the mode menu.
The polynomial is now stored in the instrument and the
method can be directly selected either by entering the
method number or else via the method-selection list.
Tipp
Save all data relating to a specific user concentration in written
form, since in the event of a loss of power (e.g. when changing
the battery) all polynomial data are lost and must be entered
anew.
The instrument has a feature enabling the user to transfer meas-
urement data to a PC via mode 67 (see section 5.6.4).
choose resolution:
1: 1
2: 0.1
3: 0.01
4: 0.001
Stored!
Chapter 5.6 | Page 10 of 13