5.5
Software update
via the internet
Before running the update
Save your stored measurement results by printing them out
or by transferring them to your computer.
When running the update these data as well as the existing
software will be entirely deleted!
Set the baud rate of the colorimeter to 19200
(mode menu, keys [Mode] [2] [9];
press [Enter],
then the [2] key,
and then use the [] or [] arrow keys
to select the baud rate).
To run the update you require:
• a PC with a Windows operating system;
• the data-transfer cable supplied with the unit;
• the supplied screwdriver; and
• the files:
Download the files as necessary and save them together in
a new folder that you have specially created for the update of
the colorimeter. You do not need to install the HexLoad.exe
programme, a simple copy is sufficient.
Please read the update instructions thoroughly before you start
to run the update.
Follow the instructions given in the update file while
performing the update.
Note
In the case of the Spectroquant® Multy Colorimeter an update
always involves a method and/or programme update.
• the programme HexLoad.exe, which is
executed on the PC and transfers the update software
to the photometer;
(see CD, section 5.5, "Software update via the
internet" or else go to
www.service-test-kits.com) on the internet; and
• the software update for the Spectroquant® Multy
Colorimeter (= *.hex file, see at
www.service-test-kits.com on the internet).
Chapter 5.5 | Page 1 of 5