Read Condition Registers; Condition; Select Default Conditions; Preset - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

Models 2500 and 2502 User's Manual

Read condition registers

CONDition?

:STATus:MEASurement:CONDition?
:STATus:QUEStionable:CONDition?
:STATus:OPERation:CONDition?
Description

Select default conditions

PRESet

:STATus:PRESet
Description
NOTE

Error queue

[:NEXT]?

:STATus:QUEue[:NEXT]?
Description
NOTE

CLEar

:STATus:QUEue:CLEar
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
These query commands are used to read the contents of the condition
registers.
When this command is sent, the following SCPI event registers are
cleared to zero (0):
1. Operation Event Enable Register.
2. Event Enable Register.
3. Measurement Event Enable Register.
The Standard Event Register is not affected by this command.
As error and status messages occur, they are placed into the error queue.
This query command is used to read those messages. See
for a list of messages.
The :STATus:QUEue[:NEXT]? query command performs the same function as
the :SYSTem:ERRor? query command. See
This action command is used to clear the error queue of messages.
SCPI Command Reference
Read Measurement Condition
Read Questionable Register
Read Operation Condition
Return registers to default conditions
Read error queue
Appendix B
"SYSTem subsystem," page 17-81.
Clear error queue
17-79