Models 2500 and 2502 User's Manual

CALCulate3 and CALCulate4

The CALC3 and CALC4 commands control relative for channel 1 and channel 2 respec-
tively. They perform the remote equivalents to the front panel REL and CONFIG REL
configuration menu.

Select input path

FEED <name>
:CALCulate3:FEED <name>
Parameters
Query
Description
FEED <name>
:CALCulate4:FEED <name>
Parameters
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
<name> = CALCulate[1]
SENSe[1]
:FEED?
This command is used to select the input path for channel 1 REL. With
CALCulate[1] selected, the REL feed will be the result of the channel 1
math function (CALC1). With SENSe[1] selected, the REL feed will be
the raw channel 1 current reading.
<name> = CALCulate2
SENSe2
:FEED?
This command is used to select the input path for channel 2 REL. With
CALCulate2 selected, the REL feed will be the result of the channel 2
math function (CALC2). With SENSe2 selected, the REL feed will be
the raw channel 2 current reading.
SCPI Command Reference
Configure and control relative functions
Select input path for channel 1 REL
Use result of channel 1 math function
Use channel 1 current reading
Query input path for channel 1 REL
Select input path for channel 2 REL
Use result of channel 2 math function
Use channel 2 current reading
Query input path for REL