Remote Ratio And Delta - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

Models 2500 and 2502 User's Manual

Remote RATIO and DELTA

Table 7-7
See the :CALC5 and :CALC6 subsystems in
Table 7-7
RATIO and DELTA function commands
:CALCulate5:DATA?
:CALCulate5:FORMat <name>
:CALCulate5:STATe <state>
:CALCulate6:DATA?
:CALCulate6:FORMat <name>
:CALCulate6:STATe <state>
:FORMat:ELEMents:CALCulate <name>
:INIT
*CALC1 (Ch. 1 math) feeds into CALC3 (Ch. 1 rel). CALC2 (Ch. 2 math) feeds into CALC4 (Ch. 2 rel). See
CALCulate subsystems in Section 17 for details.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
summarizes commands to control the RATIO and DELTA function via remote.
Command
Relative, Math, Ratio, and Delta
Section 17
for detailed information.
Description*
Request ratio reading.
Set ratio function.
Name = C3C4 (CALC3/CALC4)
C4C3 (CALC4/CALC3)
Enable/enable ratio. State = ON or OFF.
Request delta reading.
Set delta function.
Name = C3C4 (CALC3-CALC4)
C4C3 (CALC4-CALC3)
Enable/enable delta. State = ON or OFF.
Define CALC data elements. Name =
CALCulate, TIME, or STATus.
Trigger CALC readings (must use before
sending CALCn:DATA?).
7-11