Rear panel summary

The rear panel of the Model 2500 is shown in
Model 2502 is shown in
Figure 1-2
Model 2500 rear panel
VOLTAGE SOURCE
OUTPUT CHANNEL 1
INPUT connectors:
INPUT CHANNEL 1
INPUT CHANNEL 2
Triax connector terminals:
VOLTAGE SOURCE connectors:
OUTPUT CHANNEL 1
OUTPUT CHANNEL 2
WARNING
CAUTION
Digital input/output port:
DIGITAL I/O
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Figure
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
RATINGS MAX
100V @ 20mA
COMMON
MADE IN
MODE
U.S.A.
200V
INPUT
INPUT
CHANNEL 1
CHANNEL 2
!
TRIGGER LINK
I
CAT
RATINGS MAX
100V @ 20mA
VOLTAGE SOURCE
OUTPUT CHANNEL 2
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD, REPLACE FUSE WITH SAME TYPE AND RATING
Maximum common-mode voltage (voltage between INPUT common
and chassis ground) is 200V. Exceeding this value may result in a shock
hazard.
Maximum INPUT and OUTPUT connector rating is 100V @ 20mA.
Exceeding these values may result in instrument damage.
Model 2500 and 2502 User's Manual
Figure 1-2
1-3.
DIGITAL I/O
RS-232
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
100 VAC
(SB)
120 VAC
!
315 mAT
220 VAC
(SB)
240 VAC
Model 2500
Use to measure on channel 1.
Use to measure on channel 2.
Center conductor – Input HI.
Inner shell – Floating input/output common.
Outer shell – Chassis ground.
Use to output voltage bias on channel 1.
Use to output voltage bias on channel 2.
Male DB-9 connector for digital output lines, output enable line, and
component handler signals.
and the rear panel of the
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
.