Mode <Name; Clear Test Results; [:Immediate]; Auto <B - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

17-38
SCPI Command Reference
MODE <name>
:CALCulate7:CLIMits:MODE <name>
Parameters
Query
Description

Clear test results

[:IMMediate]

:CALCulate7:CLIMits:CLEar[:IMMediate]
Description
AUTO <b>
:CALCulate7:CLIMits:CLEar:AUTO <b>
Parameters
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
<name> =
GRADing
SORTing
:MODE?
This command controls how limit calculations drive the digital I/O
lines. In GRADing mode, a reading passes if it is within all of the hi/low
limit tolerances enabled, assuming that it has passed Limit 1 or Limit 2
compliance test first. The digital I/O lines will be driven with the first
pattern of the first compliance, hi, or low failure. Otherwise, the
CALC7:CLIM:PASS:SOUR3 pattern will be output.
In SORTing mode, a reading will fail if it fails the compliance test, or is
not within any of the digital I/O bands. If the tests pass and only Limit 1
or Limit 2 is enabled, the CALC7:CLIM:PASS:SOUR7 pattern will be
output. Otherwise, the first limit test band that passes will output its
LOW:SOUR3 pattern (UPP:SOUR3 patterns will be ignored). If Limit 1
or Limit 2 fails, its SOUR3 pattern will be output. If no Limit 3-6 limit
passes, the CALC7:CLIM:FAIL:SOUR3 pattern will be output.
This command clears the test results (pass or fail) of the limit tests and
resets the output lines of the digital I/O port back to the :SOURce3:TTL
settings. See
"SOURce3," page 17-74.
<b> =
1 or ON
0 or OFF
:AUTO?
With auto-clear enabled, test results will clear and the output lines of the
digital I/O port will reset when the :INITiate command is sent to start a
new test sequence.
When disabled, you must use :IMMediate to perform the clear actions.
Models 2500 and 2502 User's Manual
Control digital I/O port pass/fail output
Output graded pass/fail pattern
Output sorted pass/fail pattern
Query digital I/O pass/fail mode
Clears test results and resets digital I/O
port
Control auto-clear for test results
Enable auto-clear
Disable auto-clear
Query state of auto-clear

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2502