Figure 14-6 Measurement Event Status - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

Models 2500 and 2502 User's Manual
Figure 14-6
Measurement event status
stat:meas:cond?
(B15)
stat:meas?
(B15)
OR
To Measurement
Summary Bit
(OSB) of Status
Byte Register
(Figure 14-3)
stat:meas:enab <NRF>
stat:meas:enab?
(B15)
S2C = Source 2 In Compliance
S1C = Source 1 In Compliance
OE = Output Enable Asserted
BFL = Buffer Full
BAV = Buffer Available
ROF = Reading Overflow
RAV = Reading Available
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
S2C
S1C
OE
BFL
(B14)
(B13)
(B12)
(B11)
(B10)
(B9)
BFL
OE
S2C
S1C
(B10)
(B9)
(B14)
(B13)
(B12)
(B11)
&
&
&
&
&
BFL
S2C
S1C
OE
(B9)
(B14)
(B13)
(B12)
(B11)
(B10)
BAV
ROF
RAV
LP
LFH
L4
(B8)
(B7)
(B6)
(B5)
(B4)
(B3)
BAV
ROF
RAV
LP
LFH
L4
(B8)
(B7)
(B6)
(B5)
(B4)
(B3)
&
&
&
&
&
&
BAV
ROF
RAV
LP
LFH
L4
(B8)
(B7)
(B6)
(B5)
(B4)
(B3)
LP
= Limits Pass
LFH = Limit Failed High
L4
= Limit 4 Summary
L3
= Limit 3 Summary
L2
= Limit 2 Summary
L1
= Limit 1 Summary
&
= Logical AND
OR = Logical OR
Status Structure
14-15
L3
L2
L1
Measurement
(B2)
(B1)
(B0)
Condition Register
Measurement Event
L3
L2
L1
(B2)
(B1)
(B0)
Register
&
&
&
L3
L2
L1
Measurement Event
(B2)
(B1)
(B0)
Enable Register

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2502

Table of Contents