17-32
SCPI Command Reference

Read limits data

DATA?

:CALCulate7:DATA?
Description

LATest?

:CALCulate7:DATA:LATest?
Description
Configure and control limit tests
COMPliance:FAIL <name>
:CALCulate7:LIMit[1]:COMPliance:FAIL <name>
:CALCulate7:LIMit2:COMPliance:FAIL <name>
Parameters
Query
Description
[:DATA] <n>
:CALCulate7:LIMitx:LOWer[:DATA] <n>
:CALCulate7:LIMitx:UPPer[:DATA] <n>
Parameters
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
This command is used to acquire all the readings used for the CALC7
limit tests. At least one of the limit tests must be enabled to acquire limit
test readings. See
"Configure and control limit tests;" "STATe <b>,"
page 17-35.
This command works exactly like CALC7:DATA?, except it returns
only the latest limit result.
<name> =
IN
OUT
:FAIL?
These commands are used to specify the condition that will cause the
Limit 1 and Limit 2 tests to fail. With IN specified, the test will fail
when the Model 2500 goes into compliance (fixed 20mA current limit
for each voltage source). With OUT specified, the test will fail when the
Model 2500 comes out of compliance.
<n> =
-9.999999e20 to Specify limit value
9.999999e20
DEFault
MINimum
MAXimum
Models 2500 and 2502 User's Manual
Read limits data
Read latest CALC7 data
Set channel 1 compliance fail mode
Set channel 2 compliance fail mode
Fail Limit 1/2 test when unit goes into
compliance
Fail Limit 1/2 test when unit comes out
of compliance
Query when Limit 1/2 test failure occurs
Specify lower Limit x (x = 3-6)
Specify upper Limit x (x = 3-6)
Set specified lower limit to -1
Set specified upper limit to 1
Set specified limit to -9.999999e20
Set specified limit to +9.999999e20