16-2
SCPI Signal-Oriented Measurement Commands
Command summary
The signal-oriented measurement commands are used to acquire readings. You can use
these high-level instructions to control the measurement process. These commands are
summarized in
Table 16-1
Signal-oriented measurement command summary
:CONFigure:CURRent[:DC]
:CONFigure?
:FETCh?
:READ?
:MEASure[:CURRent[:DC]]?
Configuring measurement function
CONFigure:CURRent[:DC]
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Table
16-1.
Command
Configures Model 2500 for current measurements. Out-
Returns active function (amps).
Requests latest readings.
Performs an :INITiate and a :FETCh?.
One-shot measurement mode. Performs a :CONFigure
:CONFigure?
This command configures the instrument to measure DC current. The
:READ? command is then typically used to trigger a specified number
of measurements. See
When this command is sent, the Model 2500 will be configured as
follows:
•
Select current function.
•
All controls related to the current function are defaulted to the
*RST values.
•
The event control source of the Trigger Model is set to Immediate.
•
The count values of the Trigger Model are set to one.
•
The Delay of the Trigger Model is set to zero.
•
All math calculations are disabled.
•
Buffer operation is disabled.
•
Auto zero is enabled.
•
Both source outputs will turn on.
Models 2500 and 2502 User's Manual
Description
put turns on.
and a :READ?.
Returns active function (current).
"READ?," page
16-4.