16-2
SCPI Signal-Oriented Measurement Commands

Command summary

The signal-oriented measurement commands are used to acquire readings. You can use
these high-level instructions to control the measurement process. These commands are
summarized in
Table 16-1
Signal-oriented measurement command summary

:CONFigure:CURRent[:DC]

:CONFigure?
:FETCh?
:READ?
:MEASure[:CURRent[:DC]]?
Configuring measurement function
CONFigure:CURRent[:DC]
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Table
16-1.
Command
Configures Model 2500 for current measurements. Out-
Returns active function (amps).
Requests latest readings.
Performs an :INITiate and a :FETCh?.
One-shot measurement mode. Performs a :CONFigure
:CONFigure?
This command configures the instrument to measure DC current. The
:READ? command is then typically used to trigger a specified number
of measurements. See
When this command is sent, the Model 2500 will be configured as
follows:
Select current function.
All controls related to the current function are defaulted to the
*RST values.
The event control source of the Trigger Model is set to Immediate.
The count values of the Trigger Model are set to one.
The Delay of the Trigger Model is set to zero.
All math calculations are disabled.
Buffer operation is disabled.
Auto zero is enabled.
Both source outputs will turn on.
Models 2500 and 2502 User's Manual
Description
put turns on.
and a :READ?.
Returns active function (current).
"READ?," page
16-4.