Filter Programming Example - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

6-14
Range, Digits, Speed, and Filters

Filter programming example

Table 6-7
follows:
• Average filter off
• Median filter state: on
• Median filter rank: 5
• Average filter state: on
• Average filter type: moving
• Average filter count 20
• Advanced filter state: on
• Advance filter tolerance: 10%
Table 6-7
Filter programming example
:MED:RANK 5
:MED ON
:AVER:TCON MOV
:AVER:COUN 20
:AVER ON
:AVER:ADV OFF
:AVER:ADV:NTOL 10
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
summarizes the command sequence to program channel 1 filter aspects as
Command
Set median rank to 5.
Enable median filter.
Select moving filter type.
Set moving count to 20.
Enable average filter.
Disable advanced filter.
Set filter tolerance to 10%.