OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OUTPut <event list>Arm layer event
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut <event list>
Parameters
NOTE
Query
Description
NOTE
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Arm Layer Trigger
<event list> = TRIGger
NONE
Trigger Layer Triggers
<event list> = SOURce
DELay
SENSe
NONE
Each event in the list must be separated by a comma (i.e., :arm:output source,
delay, sense).
:OUTPut?
This command is used to specify when trigger pulses occur on the spec-
ified output trigger line of the Trigger Link (:OLINe).
Arm Layer Trigger — With TRIGger selected, an output trigger will
occur when exiting the trigger layer. With NONE selected, the arm layer
output trigger is disabled.
Trigger Layer Triggers — You can specify from one to all three events.
Each event in the list must be separated by a comma (,).
The SOURce, DELay and MEASure events refer to the Source-Delay-
Measure (SDM) cycle. This is the Device Action in the Trigger Model.
See
Figure 10-7
trigger occurs after the source is set. With DELay specified, an output
trigger occurs after the delay period. With MEASure specified, an out-
put trigger occurs after the measurement.
When disabling triggers, the NONE parameter must be sent alone (i.e., trig:outp
none). If it is listed with any of the other event parameters, NONE will be
ignored.
Models 2500 and 2502 User's Manual
Trigger layer events
Trigger on exiting trigger layer
Disable arm layer output trigger
Output trigger after source level is set
Output trigger after delay period
Output Trigger after measurement
Disable trigger layer triggers
Query output trigger event(s)
and