Operation Event Register; Figure 14-5 Operation Event Status - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

Models 2500 and 2502 User's Manual

Operation event register

The used bits of the operation event register
• Bit B0, Calibrating (Cal) — Set bit indicates that the Model 2500 is calibrating.
• Bits B1 through B4 — Not used.
• Bit B5, Waiting for Trigger Event (Trig) — Set bit indicates that the Model 2500 is
in the trigger layer waiting for a TLINK trigger event to occur.
• Bit B6, Waiting for Arm Event (Arm) — Set bit indicates that the Model 2500 is in
the arm layer waiting for an arm event to occur.
• Bits B7 through B9 — Not used.
• Bit B10, Idle State (Idle) — Set bit indicates the Model 2500 is in the idle state.
• Bits B11 through B15 — Not used.
Figure 14-5
Operation event status
Stat: oper: cond ?
Stat: oper: ?
OR
To Operation
Summary Bit
(OSB) of Status
Byte Register
(Figure 14-3)
:stat: oper: enab <NRf>
:stat: oper: enab ?
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Idle
(B9 - B7)
(B15 - B11)
(B10)
Idle
(B9 - B7)
(B15 - B11)
(B10)
&
Idle
(B9 - B7)
(B15 - B11)
(B10)
Idle = In idle state
Arm = Waiting for arm event
Trg = Waiting for trigger event
Swp = Sweeping
Cal = Calibrating
(Figure
14-5) are described as follows:
Arm
Trig
Swp
(B6)
(B5)
(B4)
(B3)
Arm
Trig
Swp
(B6)
(B5)
(B4)
(B3)
&
&
&
Arm
Trig
Swp
(B6)
(B5)
(B4)
(B3)
& = Logical AND
OR = Logical OR
Status Structure
14-13
Operation
Cal
Condition
(B0)
(B2)
(B1)
Register
Operation
Cal
Event
(B2)
(B1)
(B0)
Register
&
Operation
Event
Cal
Enable
(B2)
(B1)
(B0)
Register

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2502

Table of Contents