Front Panel Source Delay; Remote Command Source Delay - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

Models 2500 and 2502 User's Manual

Front panel source delay

To set the manual source delay from the front panel:
1. Press CONFIG then SRC1 or SRC2.
2. Select DELAY from the displayed choices, then press ENTER.
3. Enter the desired DELAY value, then press ENTER.
4. Press EXIT to return to normal display.
To set the auto source delay state from the front panel:
1. Press CONFIG then SRC1 or SRC2.
2. Select AUTO-DELAY from the displayed choices, then press ENTER.
3. Select ENABLE or DISABLE as desired, then press ENTER.
4. Press EXIT to return to normal display.
NOTE

Remote command source delay

Use the appropriate command shown in
(See
Section 17
50ms:
:SOUR1:DEL 0.05
Table 3-4
Source delay commands
:SOURce[1]:DELay <Delay>
:SOURce[1]:DELay:AUTO <State>
:SOURce2:DELay <Delay>
:SOURce2:DELay:AUTO <State>
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
The delay period is the same for both sources. If you change the delay on one
channel, the delay period on the other source will change to the same value.
for details.) For example, the following command sets the source delay to
Command
Table 3-4
to program the source delay via remote.
Description
Set source delay.
Enable/disable auto delay (ON or OFF).
Set source delay.
Enable/disable auto delay (ON or OFF).
Basic Operation