Model 2500 Setup - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

Models 2500 and 2502 User's Manual

Model 2500 setup

Step 1. Restore bench defaults.
Press the MENU key, select SAVESETUP, then press the ENTER key. From the
SAVESETUP menu, select RESET, then press ENTER. Select BENCH, then press
ENTER.
Step 2. Configure channel.
Press MSR1 or MSR2 key to select the channel to measure, then select the desired range.
Press SRC1 or SRC2 to select the voltage source, then set the source to the desired value
using the EDIT or numeric keys.
Step 3. Set up trigger parameters.
Press the CONFIG key, and then the TRIG key to access the CONFIGURE TRIGGER
menu. Select TRIG-LAYER and then press ENTER to access the CONFIGURE TRIG-
LAYER menu.
Step 4. Set trigger-in event to TRIGGER-LINK.
Select TRIGGER-IN, then press ENTER. Select TRIGGER-LINK, then press ENTER.
Step 5. Set trigger input line to #2.
Select #2, then press ENTER three times to return to CONFIGURE TRIG-LAYER menu.
Step 6. Set trigger output line to #1.
Select TRIGGER-OUT, then press ENTER. Select LINE, then press ENTER. Select #1,
then press ENTER.
Step 7. Set trigger out events to MEAS=ON (all others to OFF).
Select EVENTS, then press ENTER. Select MEAS=OFF and toggle the value to ON
using the EDIT
FIGURE TRIG-LAYER menu.
Step 8. Set trigger count to 10.
Use EDIT
press ENTER. Set the count to 10, then press the ENTER key. Press the EXIT key twice to
leave CONFIGURE TRIGGER menus.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
and
keys. Press ENTER, and then press EXIT to return to the CON-
key to scroll to the far right of the menu selections and select COUNT, then
Triggering
10-11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2502

Table of Contents