10-24
Triggering

Remote trigger commands

Table 10-1
detail in
in
Section
Table 10-1
Remote trigger command
Command
:INITiate
:ABORt
:ARM:COUNt <n>
:ARM:SOURce <name>
:ARM:TIMer <n>
:ARM:DIRection <name>
:ARM:ILINe <NRf>
:ARM:OLINe <NRf>
:ARM:OUTPut <event list>
:TRIGger:CLEar
:TRIGger:COUNt <n>
:TRIGger:DELay <n>
:TRIGger:SOURce <name>
:TRIGger:DIRection <name>
:TRIGger:ILINe <NRf>
:TRIGger:OLINe <NRf>
:TRIGger:INPut <event list>
:TRIGger:OUTPut <event list>
*TRG
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
summarizes remote trigger commands. These commands are covered in more
Section
17,
"Trigger subsystem,"
15.
Take Model 2500 out of idle state.
Abort operation, return to idle.
Set arm count (n = count).
Specify arm control source. (Name = IMMediate, TLINk, TIMer,
MANual, BUS, NSTest, PSTest, or BSTest.)
Set arm layer timer interval (n = interval).
Control arm bypass. (Name = SOURce or ACCeptor.)
Select arm layer input line (NRf = input line #).
Select arm layer output line (NRf = output line #).
Select arm layer output events. (Event list = TRIGger or NONE.)
Clear any pending input triggers immediately.
Set trigger count (n = count).
Set trigger delay (n = delay).
Specify trigger control source. (Name = IMMediate or TLINk.)
Control trigger bypass. (Name = SOURce or ACCeptor.)
Select trigger layer input line (NRf = input line).
Select trigger layer output line (NRf = output line).
Select trigger input layer events. (Event list = SOURce, DELay,
SENSe, or NONE.)
Select trigger layer output events. (Event list = SOURce, DELay,
SENSe, or NONE.)
Trigger Model 2500 (if BUS source selected).
Models 2500 and 2502 User's Manual