CALCulate6

The CALC6 commands control the delta function. They perform the remote equivalents to
the front panel DELTA and CONFIG DELTA configuration menu.

Select DELTA calculation mode

FORMat <name>
:CALCulate6:FORMat <name>
Parameters
Query
Description

Enable and read DELTA result

STATe <b>
:CALCulate6:STATe <b>
Parameters
Query
Description

DATA?

:CALCulate6:DATA?
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
<name> = C3C4
C4C3
:FORMat?
This command selects the DELTA calculation mode. With C3C4
selected, the CALC4 result (channel 2 REL) is subtracted from the
CALC3 result (channel 1 REL). With C4C3 selected, the CALC3 result
is subtracted from the CALC4 result.
<b> =
1 or ON
0 or OFF
:STATe?
This command is used to enable or disable DELTA mode. When
enabled, the CALC6 reading will be DELTA value determined by
FORMat (above). When disabled, or if a divide by zero error occurs, the
CALC6 reading will return a NAN (not a number) value of +9.91e37.
This query command is used to read the result of the CALC5 ratio cal-
culation. When enabled, the CALC5 reading will be ratio value deter-
mined by FORMat (above). When disabled, or if a divide by zero error
occurs, the CALC5 reading will return a NAN (not a number) value of
+9.91e37.
Models 2500 and 2502 User's Manual
Configure and control DELTA function
Select DELTA calculation mode
CALC3 - CALC4
CALC4 - CALC3
Query selected DELTA calculation
Control DELTA
Enable DELTA
Disable DELTA
Query state of DELTA
Read DELTA result