Models 2500 and 2502 User's Manual

CALCulate7

The following commands are used to configure and control the six limit tests:
• Limit 1: channel 1 compliance (hardware) limit test
• Limit 2: channel 2 compliance (hardware) limit test
• Limit 3 through Limit 6: software limits
When used with a handler to provide binning operations, communication between the
Model 2500 and the handler is provided via the Digital I/O port. (See
plete details.) Many control aspects of the digital output lines are performed from the
SOURce3 Subsystem. These control aspects include setting and clearing the digital output
lines, and setting pulse width. See

Select input path

FEED <name>
:CALCulate7:FEED <name>
Parameters
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
"SOURce3," page 17-74,
<name> = CALCulate3
CALCulate4
CALCulate5
CALCulate6
:FEED?
This command is used to select the input path for the Limit 3-6 tests
(software limits). With CALCulate3 or CALCulate4 selected, the speci-
fied limits will be compared to the result of CALC3 (channel 1 REL
feed) or CALC4 (channel 2 REL feed). With CALCulate5 selected, the
specified limits will be compared with the ratio result. With
CALCulate6 selected, the specified limits will be compared with the
delta result.
SCPI Command Reference
Configure and control limit tests
for details.
Select input path for Limit 3-6 tests
Use CALC3 result (channel 1 REL)
Use CALC4 result (channel 2 REL)
Use CALC5 result (ratio)
Use CALC6 result (delta)
Query input path for Limit 3-6 tests
17-31
Section 11
for com-