Input <Event List - Keithley 2500 User Manual

Photodiode meter
Hide thumbs Also See for 2500:
Table of Contents

Advertisement

17-94
SCPI Command Reference
INPut <event list>
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut <event list>
Parameters
NOTE
Query
Description
NOTE
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
<event list> = SOURce
DELay
SENSe
NONE
Each event in the list must be separated by a comma (i.e. trigger:input source,
delay, sense).
:INPut?
When TLINk is the selected Trigger Layer control source, and an event
detector in the Trigger Layer is enabled, operation will hold up at that
detector until an input trigger is received via the Trigger Link. When the
event detector is disabled, operation will not hold up. It continues and
performs the appropriate action.
A Trigger Layer event detector is enabled by including the parameter
name in the event list for the INPut command. For example, to enable
the Source Event Detector and Measure Event Detector, send the fol-
lowing command:
:TRIGger:INPut SOURce, SENSe
The Delay Event Detector will be disabled since the DELay parameter
is not included in the above event list.
To disable all the Trigger Layer event detectors, the NONE parameter must be
sent alone (i.e. trigger:input none). If it is listed with any of the other parame-
ters, NONE will be ignored.
Models 2500 and 2502 User's Manual
Enable event detectors
Enable Source Event Detector
Enable Delay Event Detector
Enable Measure Event Detector
Disable all event detectors in Trigger
Layer
Query enabled event detectors in Trigger
Layer