Models 2500 and 2502 User's Manual
STEP <n>
:SOURce[1]:VOLTage:STEP <n>
:SOURce2:VOLTage:STEP <n>
Parameters
Query
Description
NOTE
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
For example, assume you are testing a device that operates at 10V, and
you want to sweep from 8 to 12 volts. To do this, you would specify the
center to be 10V and the span to be 4 volts (12 - 8).
Use the :STEP or :POINts command to specify the number of source
and measure points in the sweep.
:CENTer and :SPAN are coupled to :STARt and :STOP. Thus, when
center and span values are changed, the values for start and stop are
affected as follows:
Start = Center - (Span / 2)
Stop = Center + (Span / 2)
<n> =
-200 to 200
DEFault
MINimum
MAXimum
:STEP
:STEP? DEFault
:STEP? MINimum
:STEP? MAXimum
These commands are used to specify a step size for a linear sweep.
When the sweep is started, the source level changes in equal steps from
the start level to the stop level. A measurement is performed at each
source step (including the start and stop levels).
This command cannot be used for a logarithmic sweep. Use the :POINts com-
mand to set the source and measure points for a log sweep.
To avoid a setting conflicts error, make sure the step size is greater than
the start value and less than the stop value.
The number of source and measure points in a linear sweep can be cal-
culated as follows:
Points = [(Stop - Start) / Step] + 1
Points = (Span / Step) + 1
An alternate way to set the source and measure points in a linear sweep
is to simply specify the number of source and measure points in the
sweep using the :POINts command.
Note that the :STEP and :POINts commands are coupled. Changing the
step size also changes the number of source and measure points. Con-
versely, changing the number of source and measure points changes the
step size.
SCPI Command Reference
Specify step size (source 1 sweep)
Specify step size (source 2 sweep)
Set source level (volts)
0V
-200V
+200V
Query step size for sweep
Query *RST default level
Query lowest allowable level
Query highest allowable level
17-69