To Install The Optional Tcd Capillary Column Adapter - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

12
Maintaining the TCD

To Install the Optional TCD Capillary Column Adapter

C A U T I O N
236
1
Gather the following:
Capillary column adapter (See
for the TCD"
on page 232.)
1/4- inch and 7/16- inch wrenches
1/8- inch brass nut
1/8- inch Vespel ferrule
Lint- free gloves
Load the
GC maintenance method
2
become ready.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble the brass nut and ferrule onto the capillary
3
column adapter.
Ferrule, 1/8-inch
Nut, 1/8-inch
Capillary column adapter
Install the adapter assembly into the detector fitting and
4
tighten finger- tight. Use a wrench to tighten until snug.
"Consumables and Parts
and wait for the GC to
Maintaining Your GC

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents