Troubleshooting and Diagnostics
Troubleshooting Guide for the Sample Transport Assembly
Troubleshooting Guide for the Sample Transport Assembly
This troubleshooting guide is meant to help you diagnose and repair
autosampler problems.
In general, autosampler problems can be divided into three categories.
1 Intermittent lock-ups with or without vial in the gripper fingers with error
2 Jittery (shaky) movement in X and/or theta axes and/or when the needle
3 Poor alignment, seen during vial pickup and vial replacement and/or when
Motor 0=X; 1=Z; 2=Theta; 3=Gripper.
N O T E
122
messages
Many times the sampler is being used very heavily.
•
motor overtemp (0 or 1 or 2 or 3)
•
movement failed (0 or 1 or 2 or 3)
•
missing vial
goes through the gripper arm into the vial with error messages
•
motor overtemp (0 or 2)
•
movement failed (0 or 2)
the needle hits the gripper arm with error messages
•
motor overtemp (0 or 2 or 3)
•
movement failed (0 or 2 or 3)
•
missing vial