Installing the Column Compartment

Installing Valve Heads (G1316C SL Plus)

Installing Valve Heads (G1316C SL Plus)

Several optional valve heads are available for the G1316C SL Plus, which can
be installed and exchanged easily.
Parts required
#
1
1
Valve Damage
C A U T I O N
Using a low pressure valve on the high pressure side can damage the valve.
Column Damage or Bias Measurement Results
C A U T I O N
Switching the valve to a wrong position can damage the column or bias measurement
results.
42
Part number
Description
5067-4107
8pos/9prt valve head high pressure and/or
5067-4108
8pos/9prt valve head low pressure
When using multiple column compartments as part of a method development
solution, make sure that the high pressure valve head is connected to the
autosampler and the low pressure valve head is connected to the detector.
For details, please refer to the Method Development Solution User and Installation
Guide (
part number: G4230-90000
Fitting the lobe to the groove is essential for making sure the valve is switched to
the correct position.
).
1200 Series TCC User Manual