Maintenance and safety inspections

Maintenance and repair of the devices (also opening the device) are to be performed only by
N
service engineers authorized and certified by h/p/cosmos, preferably within the scope of a
maintenance contract.
In case of a detected and/or assumed malfunction and/or defects and/or unreadable safety labels,
N
the device is to be disengaged immediately. Mark the device and ensure that it cannot be
operated. Inform thesupplier and authorized service personnel in writing immediately.
Disregarding warnings, notifications of intended and forbidden use, precautions as well as
N
unauthorized or lack of maintenance and/or regular safety checks may lead to injuries or death
and/or can damage the device. Furthermore it will result in loss of any liability and warranty.
Before intervenion in the device, for safety reasons, switch the running machine off and pull the
N
mains plug.
During all maintenance work and safety tests make sure that no third parties are directly or
N
indirectly in contact with the device under test and/or the technician performing the test. Keep a
safety zone of 2 m radius clear.
Some running machine models are not equipped with a UserTerminal. For these models an additional keyboard or an
external UserTerminal for remote control via RS232 is available as an option (see chapter entitled "Accessories". For
maintenance and diagnosis we recommend the use of PC software h/p/cosmos para control
Download:
www.h-p-cosmos.com/en/software/para_control.htm
8.1
Preventive maintenance
The h/p/cosmos authorised service engineers are happy to help you in case problems occur.
Preventive maintenance can prevent future problems and is indispensable for the safety of technical devices. Therefore, ask
our service department for an annual preventive maintenance contract, which is highly recommended by the manufacturer.
Some basic regular maintenance and regular safety checks as stipulated in the following chapters are obligatory!
Before switching the device on always check
the circuit cable
N
plugs
N
outlet socket
N
circuit entry of the device
N
accessories such as harnesses and cables where applicable and available
N
8.2
Immediate maintenance
Immediate maintenance is necessary if:
the device has been under high mechanical stress (push, power supply cable and/or interface cable defect through
N
driving over it or pulling it)
fluid has entered the device
N
cable and/or connector plug have been damaged
N
coverings and/or safety warnings have fallen off
N
connections made of rubber show cracks (mainly the running belt and the drive belt)
N
the running belt does not run concentrically
N
the running belt is not well-lubricated
N
a defect or malfunction of the device has been detected or is suspected.
N
Only a properly and regularly serviced device is safe.
8.3
Regular inspections / examinations
© 2012 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
maintenance and safety inspections
®
created 04.05.2012
printed 04.05.2012
page: 96 of 191