Maintenance And Safety Inspections - h/p/cosmos 150/50 Operating And Service Manual

Table of Contents

Advertisement

Do not modify the running machines and do not connect to other equipment which is not explicitly declared
N
compatible by all involved manufacturers.
Only connect accessories, software and host equipment if it is confirmed as compatible by all manufacturers.
N
It is not allowed to connect sports treadmills to medical devices such as ECG-systems.
N
Connecting a running machine to a medical device results in a medical system. Only trained staff is authorized to
N
perform this connection. Always use IEC 60601-1 approved potential isolation components. This medical system is
to be connected via a potential balance cable with the provided connector bolt and bearing within the designated
room.
The power consumption of the devices which are connected to the IEC sockets of the external UserTerminal
N
(models venus and saturn) must be in total lower than 300 W.
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the device. This may cause
N
electric shock or injury by accidental contact with moving parts.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet hand. This may cause
N
electric shock.
Never put containers with liquid on the device as this may cause electric shock or short circuit when the liquid is
N
spilled.
Never bind, process, or step on the power supply cord, or never damage or break the power supply plug. A broken
N
supply cord or plug may cause fire or electric shock.
Do not use the supply cord if its plug is loose. Such supply cord may cause fire or electric shock.
N
When removing the plug from the power supply outlet, grip the power supply plug, not the cord. Pulling the cord
N
may result in electric shock or fire by short circuit.
Disconnect the power supply plug before moving the device. Take care not to damage the power cord. A damaged
N
cord may cause electric shock or fire.
Disconnect the power plug when the unit is not used for long periods. Keeping the connection may cause electric
N
shock, current leakage, or fire due to the deterioration of insulation.
If the device is to be stored unused in an unsupervised area for an extended period, ensure that children do not
N
have access.
Do not put packing plastic bags, plastic foils or other materials within reach of children as suffocation may result.
N
7.1.3

Maintenance and safety inspections

Maintenance and repair of the devices (also opening of the device) have to be performed by service engineers
N
authorized and certified by h/p/cosmos, preferably within the scope of a maintenance contract. The installation by
unqualified personnel may cause electric shock or fire or other damages or electric shock or injury due to a
malfunction. Never disassemble, repair, or modify the device yourself.
In case of any detected and/or assumed malfunction and/or defects and/or unreadable safety labels, the device is
N
to be disengaged immediately. The device is to be marked and secured against operation and the supplier and
authorized service personnel has to be informed in writing immediately.
Before intervention in the device, switch the running machine off and pull the mains plug from the power supply.
N
During all maintenance work and safety tests, ensure that no third parties are directly or indirectly in contact with
N
the device under test and/or the technician performing the test. Keep a safety zone of 2 m radius clear.
Due to environmental influences, interior climate and type of application (for example frequent short-run operation
N
or infrequent long-run operation, etc.), the required automatic oil setting of the treadmill can deviate extensively
from the factory setting.
In case the required lubrication, quantity and quality is not assured due to lack of inspections by the user and/or
N
lack of maintenance by authorized h/p/cosmos service technicians, early damage to the running belt and running
surface can occur and warranty will expire.
In case of any maintenance work on the treadmill, ties have to be removed. The tie might be pulled in and the
N
person wearing will be strangled
© 2013 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
author: fh
email@h-p-cosmos.com
safety, installation and commissioning
created 07.02.2013
printed 04.03.2013
page: 81 of 197