Storage - h/p/cosmos 150/50 mercury Operating And Service Manual

Running machine series
Table of Contents

Advertisement

Transport of oversize running machines:
Large machines arebest moved through the window with the help of a special basket and a mobile crane.
Transports of heavy devices must only be carried out by authorised staff in compliance with the
safety standards. Otherwise there is serious danger for people and devices.
7.4.3

Storage

Store the devices at a temperature of – 20° ... + 50° C. All devices can be stored without power connection and without
operation for a period of 6 – 9 months. After this period, batteries in the device may be discharged. Maintenance and/or
a new setup/programming of the device by an authorized h/p/cosmos technician may be necessary
7.4.4
Environmental requirements / room configuration
The bearing capacity of floor and ceiling in the building must be higher than the weight of the machine. An authorized
body of the operator must certify the bearing capacity for the h/pcosmos device. For example, the h/p/cosmos saturn
300/100r: dimensions ground frame running machine: L: 3400 mm x B: 1490 mm = 5.066 m² platform. Net static weight
of running machine: 1300 kg, static body weight of subject: 200 kg, dynamical weight of subject: 1200 kg (up to 6 times
the body weight), total weight load of system at platform: 2500 kg = required dynamic bearing capacity (493.48 kg (~
500 kg)/m²). Note: Usually the bearing capacity data of the building are indicated in the static load, so that static load of
the building and static load of the treadmill, including subject should be compared.
Running machines are not to be used in medically utilized rooms with a danger of explosions or in easily inflammable
atmospheres. The devices must not be installed near to e.g. an x-ray device, motors or transformers with high voltage
connection, as the electric and magnetic interference can falsify measurements or even render them impossible. High
voltage lines must be avoided nearby the device. h/p/cosmos electrical devices with mains connections must neither be
used in wet and humid areas (e.g. swimming pools, saunas, etc.) nor in environmental chambers. If not stated otherwise
in the delivery information, h/p/cosmos devices are designed for operation in normal climatic surroundings:
Temperature:
Relative humidity:
Air pressure:
Maximum operating altitude:
The running machine should be protected from high humidity. Vents are not to be covered, as this would hinder the air
circulation.
© 2012 h/p/cosmos sports & medical gmbh
+ 10° ... + 40°C
30 ... 70 % (non condensing!)
700 ... 1060 mbar
approx. 10,000 feet (3000 m), without pressurization
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
safety, installation and commissioning
created 04.05.2012
printed 04.05.2012
page: 78 of 191

Advertisement

Table of Contents
loading

Table of Contents