h/p/cosmos 150/50 mercury Operating And Service Manual

h/p/cosmos 150/50 mercury Operating And Service Manual

Running machine series
Table of Contents

Advertisement

h/p/cosmos
model names
®
h/p/cosmos mercury
/ mercury
®
h/p/cosmos stratos
/ stratos
®
h/p/cosmos locomotion
150/50 E / DE med
®
h/p/cosmos Kistler Gaitway
incl. lt-models
h/p/cosmos quasar
/ quasar
®
h/p/cosmos stellar
/ stellar
®
h/p/cosmos pulsar
/ pulsar
®
incl. lt-models
h/p/cosmos venus
- all sizes
®
h/p/cosmos saturn
- all sizes
®
This manual is only valid for the firmware-/software version noted on the first page of this manual and only for the original
configuration of the first delivery of the machine. Firmware updates, software updates, changes of the system configuration
or retrofittings of additional equipment or accessories can result in invalidity of this manual. In case of alterations of the
device or the additional equipment; the latest version of the manual or the corresponding additional information always
should be considered.
© 2012 h/p/cosmos sports & medical gmbh
operating and service manual
h/p/cosmos
running machine series
150/50 (mercury
N
170-190/65 (quasar
N
200-450/75-300 (venus
N
med
®
med
®
II F / S
®
med
®
med
®
3p
®
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
author: mz
)
®
/ pulsar
)
®
®
/ saturn
®
development, production, sales & service
h/p/cosmos
Am Sportplatz 8
DE 83365 Nussdorf-Traunstein
Germany
phone +49 86 69 86 42 0
fax +49 86 69 86 42 49
service@h-p-cosmos.com
www.h-p-cosmos.com
manual-version
firmware MCU5-v1.06, Revision 20120502
manual order-no.: [cos14310m5-v1.06hpc-en]
email@h-p-cosmos.com
created 04.05.2012
)
®
sports & medical gmbh
®
printed 04.05.2012
page: 1 of 191

Advertisement

Table of Contents
loading

Summary of Contents for h/p/cosmos 150/50 mercury

  • Page 1 In case of alterations of the device or the additional equipment; the latest version of the manual or the corresponding additional information always should be considered. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: mz email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 2 © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 2 of 191...
  • Page 3 All kinds of installation and repair work and most maintenance work are to be performed only by trained and authorized technicians who have been certified by h/p/cosmos. The followig symbols will indicate which work can be done by the customer and which work has to be done only by authorized technicians: The customer / user should perform this maintenance and monitoring work.
  • Page 4: Table Of Contents

    Standard settings for the user options ....................... 70     Operating manual: running machines for sports and fitness c ................71     file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: mz email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 4 of 191...
  • Page 5     8.3.1 Visual inspections ............................ 97     8.3.2 Protective earth resistance (R ) measurement ..................98     file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 5 of 191...
  • Page 6 Cleansing and adjustment of the speed sensor / light barrier ................110     8.12 Light barrier at belt re-entry zones ..........................111     file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 6 of 191...
  • Page 7   12.9 Vests for subjects for airwalk systems.......................... 154     12.10 h/p/cosmos robowalk expander ..........................154 ®     file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 7 of 191...
  • Page 8 5.0 – page 2 ................ 189   ® ®   Appendix 3: Symbols ..............................190     Contact ..................................191     file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 8 of 191...
  • Page 9 300/10 ® h/p/cosmos s saturn 300/12 ® h/p/cosmos s saturn 450/30 00rs ® file: n:\ \article\cos14310m5-v1_0 06hpc-en\20120504_cos1 14310m5-v1.06hpc-en_m manual_h-p-cosmos_runn ning_machine.docx © 2 012 h/p/cosmos sports & & medical gmbh author: email@h-p-cosmos s.com created 04.05.2 printed 04.05.2012 page: 9 of 191...
  • Page 10: Model Overview

    ® h/p/cosmos saturn 300/125r h/p/cosmos saturn ® 450/300rs file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 10 of 191...
  • Page 11: Introduction

    A running machine (also called “treadmill ergometer”) is an exercise device for running or walking while staying in one place. The h/p/cosmos running machines are stationary treadmills which are used for training and testing, simulating various adjustable speeds and various adjustable elevations (also called “slope”). The medical versions can also be used for diagnostics and therapy applications.
  • Page 12: Safety Equipment

    Unweighting systems as an accessory (for example the h/p/cosmos airwalk) can be used for body weight support of the patient, in case the patient should not bear his full body weight on his legs.
  • Page 13 \article\cos14310m5-v1_0 06hpc-en\20120504_cos1 14310m5-v1.06hpc-en_m manual_h-p-cosmos_runn ning_machine.docx © 20 12 h/p/cosmos sports & m medical gmbh author: m email@h-p-cosmos. .com created 04.05.20 printed 04.05.2012 page: 13 of 191...
  • Page 14: Intended Use, Contraindications, Risks & Forbidden Use

    Intended use, contraindications, risks & forbidden use The intended application, discussed in the following, can be performed with any h/p/cosmos treadmill model, although there are more or less recommended solutions. Consider obligatory safety accessories (for example unweighting systems, optional safety arch with chest belt system for fall stop prevention and optional arm support for additional stabilization of the patient, adjustable handrails, etc.) as described in the respective chapters.
  • Page 15: Contraindications

    American College of Cardiology Foundation - www.acc.org American Heart Association - www.americanheart.org http://my.americanheart.org/idc/groups/ahaecc-internal/@wcm/@sop/documents/downloadable/ucm_423807.pdf Further contraindications may occur. This has to be evaluated by the responsible medical doctor. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 15 of 191...
  • Page 16: Risks

    Non-cardiac complications Intracranial haemorrhage n.s. Thromboembolic event n.s. Transient global amnesia n.s. Arthralgia n.s. Myalgia n.s. Low back pain n.s. n.s.: not specified file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 16 of 191...
  • Page 17 1.5 m. 4. Recommendation and application of weight support system and/or fall stop device, such as the h/p/cosmos safety arch with harness and automatic shut-off of the running belt. 5. Comprehensive instruction of the patient before the...
  • Page 18: Forbidden Use

    A safety arch with chest belt and fallstop is also obligatory in this case. h/p/cosmos devices must not be used in environmental conditions other than those specified in the chapter „Technical specifications“ and “Environmental conditions” (e.g. in wet and humid areas, swimming pools, sauna, environmental chambers, high-pressure-, low-pressure, altitude- and oxygen-chambers, etc.).
  • Page 19: Safety Notes, Warnings, Precautions

    Any other use than that explicitly listed in the field of application is prohibited. Use of the h/p/cosmos medical treadmills in a home environment and/or in public and/or operation by persons who are not medically trained staff is not allowed.
  • Page 20: Preparation Of The Patient / User

    Treadmills which are not designed for cycling, wheelchairs, spikes or ski sticks, must not be used with those applications as this will result in serious damage to the treadmill. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 21: During Treadmill Exercise

    Children are only allowed to use the running machine under constant supervision of a medical doctor and only when safeguarded by safety arch with chest belt system. Additionally h/p/cosmos recommends the use of adjustable handrails. When using the treadmill for load tests, a medical doctor and a defibrillator are to be within reach at any time.
  • Page 22 If the treadmill is being controlled via ECG, pay special attention to the ECG manufacturer’s manual! The summary of clinical data and the risk management of h/p/cosmos does NOT cover any measured parameters and/or vital functions which are measured and/or detected by host devices (e.g. ECG, ergospirometry devices, EMG, etc.) which are connected to the treadmill.
  • Page 23: Machine Care

    All wear and tear parts are to be replaced at least every two years (or earlier if necessary and/or earlier at first sign of wear and/or damage), all worn parts are to be replaced. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 24: Operation

    The user selects one of the operation modes (manual, profile, cardio or test mode) of the remote control firmware and selects the prescribed test/training. The software asks the user to confirm or to change default settings. Current values file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 25: Switch On Procedure

    The indicator light within the key (push button) is illuminated. If the indicator does not flash up, please check the power supply, the expulsion fuse and the emergency stop. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 26: Switch Off Procedure

    Press the emergency stop only in case of danger. The emergency stop is not to be used as a normal stop key. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 27: Emergency Stop, Stop Buttons And Functions And Safety Features

    The emergency stop button is to be pressed immediately! h/p/cosmos is in compliance with the norm EN 957-6 and offers more safety features than required, as shown in the following tables. In all below mentioned cases, all displayed values will be reset to “0”.
  • Page 28 -en\20120504_cos14310 m5-v1.06hpc-en_manua l_h-p-cosmos_running_m machine.docx © 2012 h/p/cosmos spo orts & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 ge: 28 of 191...
  • Page 29: Userterminal

    More recent computers will only have a USB interface instead of the RS232 interface. In that case, a „USB to RS232 interface adapter“ is available at h/p/cosmos under order number [cos12769]. For control via USB interface the processor must be a Pentium 1.8 GHz or higher. To use all functions described below and for maintenance and service, we recommend the PC software h/p/cosmos para control (Freeware).
  • Page 30: The Keyboard

    ... the mode (manual, profile, cardio, test) has been changed ... the display has been initiated by pressing the key once more file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 30 of 191...
  • Page 31 Weight height of subject H ..Standard configuration of the LCD displays and LEDs. LED on / LED off file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 31 of 191...
  • Page 32: Application: Ergometry / Cardiology Ecg Treadmill Stress Test

    ® ® Kistler force measurement plates. It is the only h/p/cosmos treadmill with these features. The system measures the vertical ground reaction forces and thereby collects a large amount of force and time-based gait parameters. The total force path can be displayed with the Kistler Gaitway software .
  • Page 33 The following figures show four examples of the clinical benefit of h/p/cosmos gaitway with KISTLER force plates: ® Arthritis in the upper ankle 4 weeks later Femoropatellar syndrome 4 weeks later After four weeks of gait therapy you can see a considerable improvement of gait symmetry, documented by comparison of the ground reaction forces.
  • Page 34: Modes Of Operation

    The acceleration levels are accessible by pressing the "speed-keys" + and minus several times and then holding the key. The acceleration levels are also accessible via interface and optional PC software h/p/cosmos para control® and h/p/cosmos para graphics®.
  • Page 35 Therefore the high accelerations are only accessible via special release in the optional functions, protected by access key. Permissable maximum and minimum acceleration levels can be limited. See chapter 5.10 entitled “Optional settings: User Options”. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 35 of 191...
  • Page 36: Manual Mode

    5.9.2 Manual mode The following table shows the operating steps for running the the treadmill via firmware or software h/p/cosmos para control. Operation of profile mode Initiation: Running belt is not moving. One of the modes LED is flashing: (manual, profile, cardio, test) Step no.
  • Page 37: Profile Mode

    High 2 Low 2 High 3 Low 3 High 4 Low 4 High 5 Low 5 High 6 Low 6 25 min file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 37 of 191...
  • Page 38 High 3 10.0 Low 3 High 4 10.0 Low 4 High 5 10.0 Low 5 High 6 10.0 Low 6 25 min file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 38 of 191...
  • Page 39 Low 3 10.1 High 4 13.0 Low 4 10.1 High 5 11.5 Low 5 10.1 High 6 11.5 Low 6 10.1 25 min file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 39 of 191...
  • Page 40 The following table shows the operating steps for running the the treadmill via firmware or software h/p/cosmos para control ® Operation of profile mode Initiation: Running belt is not moving. One of the modes LED is flashing: (manual, profile, cardio, test) Step no.
  • Page 41 Switch to the next or previous Switch to the next or previous mode. mode This is a perfect way to change the mode without stopping the belt. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 41 of 191...
  • Page 42: Cardio Mode

    Every 30 seconds a warning signal is audible, after 1 minute the running machine cancels the mode (at MCU5 firmware-versions < V1.03.1 or MCU4 firmware-versions < V4.04.2, file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 43 15.1 ... 30 30.1 ... 50 > 50 The following table show the operating steps to run the the treadmill via firmware or software h/p/cosmos para control ® Operation of cardio mode Initiation: Running-Belt is not moving. One of the modes LED is flashing: (manual, profile, cardio, test) Step no.
  • Page 44 Speed and elevation of the running machine is controlled automatically (see separate table). [11] Stop running belt Running belt stops. (Deceleration time can be adjusted in the optional functions) file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 44 of 191...
  • Page 45 Switch to the next or previous Switch to the next or previous mode. mode This is a perfect way to change the mode without stopping the belt. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 45 of 191...
  • Page 46: Test Mode

    Use a safety arch with chest belt system when making max. load and max. heart rate tests. The following table shows the operating steps to run the the treadmill via firmware or software h/p/cosmos para control Operation of test mode Initiation: running machine is in standby mode.
  • Page 47: Standard Ergometric Tests

    For correct calculation of the UKK fitness index, the following parameters have to be entered: gender, age, weight and height. The following table shows the operating steps to run the treadmill via firmware or software h/p/cosmos para control ®...
  • Page 48 Note the index and used time. [14] Cool-down approx. 5 min, then End of the UKK 2 km walking test stop manually. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 48 of 191...
  • Page 49 „-“ button, indicates PAUS To end pause you can either press START or „+“ – see: end pause manually. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 50 6. Balke protocol , e.g. for ECG stress test Step Duration (min) Speed (km/h) Elevation (%) 10.0 12.5 2:00 15.0 17.5 20.0 22.5 25.0 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 50 of 191...
  • Page 51 Target speed standard: 10.0 km/h; adjustable from 0 to maximum speed of the treadmill. Time for reaching target speed in seconds: standard: 10 seconds; adjustable from 0 to 99 seconds file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 51 of 191...
  • Page 52: Additional Profiles

    (km/h) level 600 m 14.5 1:30 min 600 m 15.5 1:30 min 600 m 16.5 1:30 min 600 m 17.5 1:30 min file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 52 of 191...
  • Page 53 0:45 min 300 m 21.0 0:45 min 400 m 19.0 1:00 min 500 m 17.0 1:15 min 600 m 15.0 1:30 min file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 53 of 191...
  • Page 54 0:45 min 300 m 20.0 0:45 min 300 m 20.0 0:45 min 300 m 20.0 0:45 min 300 m 20.0 0:45 min file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 54 of 191...
  • Page 55 Time (min) (km(h) level 2:00 2:00 11.0 1:00 11.0 1:00 11.0 1:00 11.0 1:00 11.0 1:00 11.0 1:00 12.0 2:00 12.0 2:00 12.0 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 55 of 191...
  • Page 56 1:00 1:00 1:00 1:00 1:00 1:00 1:00 2:00 2:00 2:00 86 ... 89 Not engaged / reserved for following extensions and up-dates file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 56 of 191...
  • Page 57 18.0 0:30 3:00 19.8 0:30 3:00 21.6 0:30 3:00 23.4 0:30 3:00 25.2 0:30 3:00 27.0 0:30 3:00 28.8 0:30 3:00 30.6 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 57 of 191...
  • Page 58 14.4 0:30 3:00 14.4 0:30 3:00 14.4 0:30 3:00 14.4 0:30 3:00 14.4 0:30 3:00 14.4 0:30 3:00 14.4 0:30 3:00 14.4 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 58 of 191...
  • Page 59 1:00 13.6 2:30 13.6 2:20 10.5 0:10 12.7 1:00 18.6 1:00 14.3 2:30 14.3 2:20 11.2 0:10 13.4 1:00 19.3 1:00 15.0 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 59 of 191...
  • Page 60: Programming Example: User-Defined Profile

    Pr. 21. Up to 40 program steps can be stored per program. If more program steps are required, ask for the optional PC sSoftware h/p/cosmos para graphics , which is able to control all functions of the ®...
  • Page 61 [17] Confirm time indicates time (seconds): 2:00 is flashing [18] Select time 30 seconds [19] Confirm time indicates elevation: 0.0 is flashing file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 61 of 191...
  • Page 62 Select acceleration level 4 [37] Confirm acceleration level indicates speed: 0.0 is flashing indicates program step: St. 5 [38] Select der Speed 6.0 km/h file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 62 of 191...
  • Page 63 End of programming. save program file for more than The profile can be started later on in test mode Pr. 21 5 seconds file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 63 of 191...
  • Page 64: Optional Settings: User Options

    Optional settings serve for example as confirmation of error notes on the display, or for the standard setting of the device: selection of RS232 interface protocol, etc. For optional settings on devices without UserTerminal, an external UserTerminal or a connected PC with the software h/p/cosmos para control is required.
  • Page 65 OP01 Reset (deleting) of error messages The required maintenance work has to be performed before deleting error messages. The h/p/cosmos service department has to be contacted before one of the service displays or error codes will be deleted! Note: This option resets exclusively the causing variable.
  • Page 66 9999 9999 = next sender will be accepted, saved and filtered (can also be set with UP and DOWN) (further settings in user OP23 and administrator OP16 are necessary) file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 66 of 191...
  • Page 67 Individual OFF = RS232 not active / no protocol / interface deactivated adaptation of protocol on request. 1 = h/p/cosmos coscom v1, v2, v3 with Baudrate 9600 bps For further information about the (standard setting COM 1 and COM 2) h/p/cosmos coscom protocol see 2 = JAEGER OXYCON (up to 1995);...
  • Page 68 RS 232. The maximum adjustable value depends on the setting of option 28. Note: If the peripheral equipment sends an acceleration and deceleration command via the h/p/cosmos coscom protocol, the selected level in option number 29 is not valid for these speed commands.
  • Page 69 For correct printout, the connected printer must be compatible with PCL printer language. For special characters the ISO 8859-1 (Latin-1) font is used. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 69 of 191...
  • Page 70: Standard Settings For The User Options

    0 = no single values, 1 ... 100 OP 53 Language settings for printer protocol English English, German, French, Spanish Portuguese, Hungarian file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 70 of 191...
  • Page 71: Operating Manual: Running Machines For Sports And Fitness C

    Safety notes, warnings, precautions All safety notes, warnings and precautions of the medical running machines have to be considered. 4.1 and 7.1)”. (Chapters file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 71 of 191...
  • Page 72: Service Manual

    All kind of installation and repair work as well as most maintenance work is to be carried out only by trained and authorized technicians who have been certified by h/p/cosmos. The following symbols will indicate which work can be done by the customer and which work has to be done by authorized technicians: The customer / user should perform this maintenance and monitoring work.
  • Page 73: Maintenance And Safety Inspections

    In case the required lubrication, quantity and quality is not assured due to lack of inspections by the user and/or lack of maintenance by authorized h/p/cosmos service technicians, early damage to the running belt and running surface can occur and warranty will expire.
  • Page 74: File: N:\Article\Cos14310M5-V1_06Hpc-En\20120504_Cos14310M5-V1.06Hpc-En_Manual_H-P-Cosmos_Running_Machine.docx

    Shie elded motor cab Freq quency inverter metal housing (for shie lding) Opto o coupler board file: n:\article e\cos14310m5-v1_06hpc- -en\20120504_cos14310 m5-v1.06hpc-en_manua l_h-p-cosmos_running_m machine.docx © 2012 h/p/cosmos spo orts & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 ge: 74 of 191...
  • Page 75: Medical Device Software Classification

    For this reason the worst case malfunction scenario has to be taken into account. h/p/cosmos classifies the internal firmware and software components as class B due to a number of risk reducing design features and measurments.
  • Page 76: Installation & Commissioning

    Within Germany, most devices are delivered and assembled by h/p/cosmos directly or by an authorised forwarder. If delivered by h/p/cosmos, the packaging will be removed and recycled.
  • Page 77 Trolley with 4 wheels for max. 300 kg (660 lbs) [cos13016] Trolley with 4 wheels for max. 500 kg (1100 lbs) [cos13672] file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 78: Storage

    The bearing capacity of floor and ceiling in the building must be higher than the weight of the machine. An authorized body of the operator must certify the bearing capacity for the h/pcosmos device. For example, the h/p/cosmos saturn 300/100r: dimensions ground frame running machine: L: 3400 mm x B: 1490 mm = 5.066 m² platform. Net static weight of running machine: 1300 kg, static body weight of subject: 200 kg, dynamical weight of subject: 1200 kg (up to 6 times the body weight), total weight load of system at platform: 2500 kg = required dynamic bearing capacity (493.48 kg (~...
  • Page 79 IEC 60601-1. According to this example the leakage current would flow from the electric device to the earthened subject via the trainer/doctor. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 80: Mechanical Installation

    The transportation locks have to be removed after setup on the h/p/cosmos gaitway model with integrated force plate. Drive elevation up to max of 25 %. Loosen the four M8 screws, marked with the red arrows on the underside of the aluminum force plates, and remove them.
  • Page 81: Electrical Installation

    To start again, switch on the running machine at the main ”ON“ switch and press START at the keyboard. h/p/cosmos running machines are to be installed at a voltage power connection with ground wire (earthing,) including leakage-current protection switch (interrupter) and according to VDE 0100 or/and the currently valid regulations and directives.
  • Page 82 300/125r ® ® For d details please re efer to chapter 1 11entitled "Tech hnical data”. file: n:\article e\cos14310m5-v1_06hpc- -en\20120504_cos14310 m5-v1.06hpc-en_manua l_h-p-cosmos_running_m machine.docx © 2012 h/p/cosmos spo orts & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 ge: 82 of 191...
  • Page 83: Electric Safety Measurements And "First Measured Values

    One copy of this protocol [cos11690xx] remains with the owner’s manual while the original of the protocol with the “first measured values“ should be sent to the manufacturer h/p/cosmos. Picture: Example for electric safety testing device based on IEC 601-1 7.4.6.2...
  • Page 84: Plug Assignment At The Userterminal

    If the UserTerminal was disassembled during transport, please connect the cables according to the following illustration: Devices with MCU5 control unit (as of manufacturing date 12/2007) file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 85 Devices with MCU4 control unit (until manufacturing date 12/2007) file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 85 of 191...
  • Page 86 For devices with MCU4 control unit (until manufacturing date 12/2007) and reverse belt rotation file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 86 of 191...
  • Page 87 UserTerminal Ground connection at the UserTerminal Cable connection for ON/OFF button and emergency stop in the UserTerminal. GRD = ground file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 87 of 191...
  • Page 88: Connection Of External Host And / Or Auxiliary Devices Via Interface

    All other interface protocols (older versions coscom v2, coscom v1, coscomekg.dll, other treadmill manufacturers, etc.) are not validated by h/p/cosmos according to the norms EN 62304 and EN 14971 and must not be used for medical applications (e.g. ergometry, stress-tests, cardiorespiratory diagnostics, performance diagnostics, etc.) since March 21,...
  • Page 89: Programmable, Electrical Medical Systems (Pems) / In-House Production

    In case the manufacturer declares multiple sockets as valid, these must not be located on the floor in order to prevent soaking by liquids or mechanical damage. h/p/cosmos herewith informs about all above and below mentioned risks, warnings and obligations. Responsibility for system integration: The operator of ME equipment and ME systems (ME=medical-electrical) is to appoint a so-called system integrator, responsible for taking care of the tasks that result from the norm.
  • Page 90 PACS and WEB-based electronic image distribution server-based database for long-term ECG devices or stress test systems telemedicine applications Wi-fi connectivity of patient monitors et. al. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 90 of 191...
  • Page 91 Do the medical data arrive at the right place at the right destination in their complete form? May there be unforeseen changes, the user acknowledges in time? Is there a readily available documentation for all hard- and software components with all updates? file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 92 As class B networks are usually radiological networks, it is difficult to define the responsibility for such a system when several manufacturers allocate different modules. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 93 In particular, the third edition of DIN EN 60601-1 is directed much more at the operator of telemedicine connections than the second edition. In order to operate a safe and high quality tele-medicine as tele-neurology network, the safety standards should be met. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 94: Installation Checklist And Instruction Of The User

    Then, the h/p/cosmos employee / h/p/cosmos partner instructs the user according to the instruction protocol in appendix 2. It is important to include all people in the instruction and commissioning who are going to work with the running machine. After the instruction is completed, the instruction protocol has to be signed by the h/p/cosmos technician and all trained persons and returned, together with the signed delivery note and the registration form, to h/p/cosmos.
  • Page 95 Test the function of the heart rate transmission with the help of a simulator or a POLAR sender. Search for interferences and possible reasons if applicable. accomplished file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 95 of 191...
  • Page 96: Maintenance And Safety Inspections

    Preventive maintenance The h/p/cosmos authorised service engineers are happy to help you in case problems occur. Preventive maintenance can prevent future problems and is indispensable for the safety of technical devices. Therefore, ask our service department for an annual preventive maintenance contract, which is highly recommended by the manufacturer.
  • Page 97: Visual Inspections

    The main inspection sticker on the device (e.g. running machine) also certifies the inspection of the optional equipment and the accessories. However, inspection intervals for optional equipment and accessories (e.g. rope of h/p/cosmos airwalk unweighting system, chest belt for safety harness, compressors, etc.) may deviate significantly from inspection intervals of the main device.
  • Page 98 0701/0702 (sports and fitness machines) or VDE 0751/IEC 601-1 (medical devices). The measurement is equivalent to the single fault earth leakage current based on IEC 60601. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 98 of 191...
  • Page 99: Leakage Current Measurement

    In the event that the device cannot be repaired accordingly and immediately, it must be marked and secured against further operation. Informthe supplier and authorized service personnel in writing immediately. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 100: Building Installation: Electric Checks, Protective Earth Function, Rcd, Leakage Current

    An h/p/cosmos authorized technical service engineer will then advise what corrective action needs to be taken. Use original h/p/cosmos special oil only. The use of other oils or the failure to lubricate regulary may damage the running belt, running deck or the lubrication system and reduce warranty and liability. h/p/cosmos recommends a maintenance contract with an authorized h/p/cosmos service partner.
  • Page 101: Devices With Userterminal And Without Automatic Oil-Pump

    Lubricant is available on request at h/p/cosmos. The running belt should be lubricated after 1000 km at the latest and/or ifdry grinding noises are heard during operation.
  • Page 102: Lubrication Film Test

    Based on previous experience, h/p/cosmos is able to evaluate the quantity of oil absorbed by measuring the difference in weight. However, the user should realize that this test can only show approximate values and trends and is not meant to replace the direct inspection by an authorized h/p/cosmos technician at the prescribed time intervals.
  • Page 103: Devices Without Userterminal And Without Automatic Oil Pump

    Please contact h/p/cosmos to order the special oil. The running belt should be lubricated after 1000 km at the latest or/and if dry grinding noises are detected during operation. Special devices may have different intervals.
  • Page 104 For those models a lubrication test (tissue-test) according the instructions (available from h/p/cosmos) should be performed during the obligatory yearly maintenance. The test should be sent back to h/p/cosmos for evaluation.
  • Page 105: Refilling Oil Tank

    Open the lid of the tank and refill the oil-tank with approximately 0.5 litres of the provided original oil, which can be ordered at h/p/cosmos. For devices constructed before 2002 and without an external control unit, add approximately 1 to 1.5 litres) Close the lid of the oil-tank, plug in the power plug and switch the device on.
  • Page 106: Control And Tightening Of The Running Belt

    The belt tension should not be more than 0.5 %, otherwise it could damage the belt, the shafts or the bearings! allen screws for tigthening of running belt tension file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 107: Adjustment (Centring) Of The Running Belt

    The allen key must be pulled out of the screw immediately after usage due to risk of injury! The oversize machine h/p/cosmos saturn 450/300rs has an automatic belt tightening system with motors and sensors and does not need to be tightened manually.
  • Page 108: Control And Tightening Of The Driving Belt (Poly-V Belt System)

    The allen key must be pulled out of the screw immediately after use due to risk of injury! The oversize machine h/p/cosmos saturn 450/300rs has an automatic belt centering system with motors and sensors and does not need to be centered manually; however, the user should observe the central belt position.
  • Page 109: Driving Belt With Timing Belt System

    Check the anti-skid areas at regular intervals and replace them at signs of wear and tear. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 110: Hygiene And Cleansing

    The surfaces of the treadmill can be cleaned with a light moistened cloth. For disinfection h/p/cosmos recommends Bacillol plus which can be ordred from h/p/cosmos, order number [cos12179]. Before use of cleaning and disinfection substances, always test the compatibility of the substance at poorly visible places.
  • Page 111: Light Barrier At Belt Re-Entry Zones

    To avoid unwanted switch off (quick stop), the light barrier and the reflector glass should be cleaned carefully with a cloth moistened with alcohol at intervals of approximately one week, depending on the air and dust conditions in the room. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 112: Adjustment Of Light Barriers

    Spare parts & consumables order number description picture special oil (bottle 0.25 l) for h/p/cosmos running deck surface. Only for running cos000200008001 applications (not for bicycle and wheel-chair use!) and manual lubrication with "hand-pump" syringe) syringe 10 ml for lubrication of running surface cos12181 special oil (bottle 0.50 l) for automatic lubrication...
  • Page 113: Trouble Shooting

    Secondary fuse isolation transformer 1-pole B16A Secondary fuse isolation transformer 3-pole C16A Expulsion fuse on devices with external switching cabinet Fuses in cabinet for power supply unit file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 113 of 191...
  • Page 114: Electrostatic Discharge

    Pay special attention to the advice and data in the chapter 11.5 entitled “Technical data” / “ lectromagnetic compatibility (EMC) and immunity: Guidance and manufacturer's declaration”.. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 115: Voltage On The Device Housing / Electric Shock

    Do not rely on the indicated values if you suspect interference. Please consult the instructions provided by the manufacturer POLAR concerning interference at www.polar.fi or at http://www.polarservicecenter.com/Frame-Troubleshooting.htm file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 115 of 191...
  • Page 116: Rs232 Interface

    The MCU control unit also has a number of status LEDs which can assist in diagnosis. When communicating with the h/p/cosmos headquarters during error search and trouble shooting, the authorized technican should observe the indications on these LEDs and report to the h/p/cosmos headquarters.
  • Page 117: Error Messages

    Incorrect measurement of angle of elevation Maybe wrongly adjusted, dirty or defective elevation sensor Contact authorized service engineer file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 117 of 191...
  • Page 118 Check plug no. S6 on MCU and “inverter on – signal / RFR” from MCU board to inverter drive. Check mains power supply to frequency inverter. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 119 Running machine is still working. At MCU5 control electronic (as of manufacturing date 12/2007), this error shows low or defective a defect of the option memory (FRAM). In this case, please contact h/p/cosmos battery and request a service technician. Optional setup restores to default factory setting and device type.
  • Page 120 Parameter storage device fault E.PUE PU disconnection E.RET Retry count excess E.CPU CPU fault Option fault CPU fault CPU fault E.13 Internal circuit fault file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 120 of 191...
  • Page 121 Running machine is still functioning (except chip card function). Unknown chip card malfunction. No card identified. Check chip card. Insert correct chip card. Reprogram chip card. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 121 of 191...
  • Page 122 Reset of error messages on devices without UserTerminal The OIL message and SERVICE message are quit with OP01. Other error messages have to be quit by OP01 at the administrator level. In both cases, an external UserTerminal or the software h/p/cosmos para control are needed. For ®...
  • Page 123: Optional Settings: Administrator Options

    10 Optional settings: administrator options The administrator options are only accessible through h/p/cosmos staff and authorised service engineers. Access is only possible with a special code which is not given here. Changing an administrator option might cause serious problems if you are not familiar with all details. Therefore the h/p/cosmos service has to be contacted before modifying administrator options.
  • Page 124 The standard maximum speed of the selected device type (in OPTION 03) will be indicated in the SPEED display. Example: h/p/cosmos mercury 4.0 = 22.0 km/h; h-p-cosmos discovery 4.0 = 40 m/min For safety purposes (e.g. for beginners), the max. speed of every treadmill model can be limited.
  • Page 125 (OP11=1). If the running machine is not equipped with a speed measuring system, this can be retrofitted easily. To activate the motor brake, some parameters of the frequency inverter must be changed. Ask the service department of h/p/cosmos sports & medical gmbh for detailed instructions. Attention!
  • Page 126: List Of Running Machines Types

    OFF, 100 ... 5.000 km Do not change the interval without prior communication with h/p/cosmos. For the models h/p/cosmos venus and saturn, JAEGER CareFusion LE 580 CE and larger as well as COSMED T200 and larger, OP35 is not adjustable because the machines are equipped with an automatic oil pump and tank.
  • Page 127 E.02 “HELP" performed by a trained and authorised service technician. Appropriate protocols and inspection instructions are available at h/p/cosmos service. An error message "E 02" "HELP" for the TSC-service interval will appear in the display after expiration of a time period of 12 months.
  • Page 128 It should be 10 V. Make a note of the voltage in the service report (not possible with digital control via RS485)! Read the speed and make a note of the value in the h/p/cosmos service report: OP 48: _____ km/h Stop the running belt with the STOP key.
  • Page 129 (h/p/cosmos venus since 1997) 4 = elevation control: linear motor for 0 - 24% elevation, increment sensor, lift range 240 mm (h/p/cosmos mercury models since 1997) 5 = elevation control: linear motor for 0 - 28% elevation, increment sensor, lift range 300 mm (h/p/cosmos quasar 4.0 since 2002) 6 = elevation control: linear motor for 0 - 25% elevation, increment sensor, lift range 300 mm (h/p/cosmos pulsar 4.0 since 2002)
  • Page 130 (for ladder ergometer: 8.0 m/min for 8 hours) STOP after 13 hours. Adjustable settings: 0 = test profile 99 locked / de-activated 1 = test profile 99 activated file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 130 of 191...
  • Page 131 100 ... 5.000 km OP 37 Interval for service message TSC 12 months 0 … 48 month (time-dependent) for technical safety control E.02 “HELP" file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 131 of 191...
  • Page 132 0 = test 99 de-activated 0 = test profile 99 de-activated in the factory / service 1 = test profile 99 activated file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 132 of 191...
  • Page 133 2874 40.9 113.0 190 / 65 45.0 2874 46.0 125.0 110.0 0 (3p), 125.0 ( file: n:\article e\cos14310m5-v1_06hpc- -en\20120504_cos14310 m5-v1.06hpc-en_manua l_h-p-cosmos_running_m machine.docx © 2012 h/p/cosmos spo rts & medical gmbh mail@h-p-cosmos.com created 04.05.2012 rinted 04.05.2012 page e: 133 of 191...
  • Page 134: Technical Data

    6 LCD displays / 6 keys / display of mode and unit with LED. E: no Terminal (display & keyboard) “lt-models“ and “h/p/cosmos locomotion 150/50 E med“ have no UserTerminal (no display, no DE: TouchScreen keyboard). Control via interface. Power supply...
  • Page 135 When using a front-bar handrail (crossbar), the speed is limited to 5 km/h for safety reasons. Data of weight and measurements can deviate. Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & EO. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 136: Running Machines, 170-190/65 (Quasar / Pulsar)

    0.0 ... 28.0 % - 25.0 % ... + 25.0 % not applicable not applicable Definition 0.1 % 0.1 % 0.1 % file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 136 of 191...
  • Page 137 ** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
  • Page 138: Running Machines, 190/65 (Pulsar 3P)

    0.0 ... 10.0 km/h or 0.0 ... 40.0 km/h or 0.0... 45.0 km/h Angle of inclination** - 25.0 % ... + 25.0 % Definition 0.1 % file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 138 of 191...
  • Page 139 ** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
  • Page 140: Running Machines, Oversize 200-450/75-300 (Venus / Saturn)

    3000 x 3000 x 3000 x 4500 x [r] = suitable for bicycle and wheel chair 1000 1000 1250 1000 1250 3000 application file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 140 of 191...
  • Page 141 ** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx.
  • Page 142 Please pay special attention also to the other EMC related chapters of this manual: Chapters: Safety, installation, operation, POLAR heart rate monitor, troubleshooting. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 143 NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 144: Compatible Devices

    150/50 cm and 170…190/65 cm All other systems and devices which are not in the above stated list must not be connected to h/p/cosmos running machines. Please contact service@h-p-cosmos.com for the latest updated list of compatible devices and systems in writing. Only connect and/or use accessories, software and host equipment if it is approved as compatible by all manufacturers.
  • Page 145: Interface Protocols

    First connect the potential equalization cable with the corresponding potential equalization plug pin (next to the main switch on the front of the h/p/cosmos device), then connect the entire medical system to the main potential equalization terminal of the medical room and then connect the mains voltage power plug of the medical system.
  • Page 146: Arm Support

    12 Acc cessories an nd options The follow wing chapters de escribe the acc essories and op ptions available e at h/p/cosmos Only conne ect and/or use accessories, software and host equipme ent if it is appr roved as comp...
  • Page 147 acce ssories and d options Only trained staff is allowed to adju ust the arm sup pports. Only loosen the gea ar tooth locking g elements to a adjust the angl e while holding g the arm supp ports with one h hand at the hand dle or at the up...
  • Page 148: Accessories And Options

    12.3 Special handrails h/p/cosmos offers a number of optional handrails and custom-made devices on request. Details and order numbers are available at the manufacturer h/p/cosmos or at your local authorized distributor. Handrail crossbar adjustable Handrail crossbar bended...
  • Page 149 [cos1003 only available fo or the running m machines with r running surface 150/50cm (e.g g. h/p/cosmos m mercury med) ® Repro oducibility of a adjustments To ob btain reproducib bility, the individ...
  • Page 150: Handrails, Adjustable: Parallel Bars With Gas Struts And Locking Lever

    If turnng the UserTerminal to the outside position, at least one emergency stop button has to remain in reach of the subject. The use of the h/p/cosmos arm support is not possible with the adjustable bar handrails. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com...
  • Page 151 © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 151 of 191...
  • Page 152 acce ssories and d options 12.6 herapist’s seat t and foot supp port, adjustabl The bilater rally positioned and ergonomic cally formed the erapist’s seats in combination with the adjusta table foot suppo orts provide the therap pist with a back- -friendly seating g possibility whi ile performing t...
  • Page 153: Unweighting System Airwalk Se 135 (Available As Of 10/2008)

    The training phases can be extended, which results in a more effective therapy for the patient. The “h/p/cosmos airwalk se 135“ can be used for a very wide field of applications. Therapeutic applications such as functional motion therapy, rehabilitation or neurology as well as competitive sports (e.g. coordination training for athletes, etc.).
  • Page 154: Unweighting Systems Airwalk 70/160

    35/70/160 cos13752 It is possible to use the h/p/cosmos airwalk with the running machine series with track size 150/50 cm or 170/65 cm or 190/65 cm, and as well with most running machines produced by other manufacturers.
  • Page 155: Wheelchair Ramp

    Wheel chair stabilizer for deck length 200cm: [cos10227] 250cm: [cos10226] 300cm: [cos00096110031] Available for the h/p/cosmos running machine series with track size 200/75 cm (and bigger) with the mark “r” Before using the running machine for any application without a wheelchair (running, cycling, skating etc.), the wheelchair stabilizer has to be dismantled for safety reasons.
  • Page 156: Safety Arch With Chest Belt System

    The length of the rope should be adjusted so that the subject can not use the rear 30% of the running deck. Put on the chestbelt so that the h/p/cosmos logo is at the front. In order to attach the carabiner, put the vertical belt (1) below the shoulder belts (2), facing the body, thereby unloading the seam at the joint (3).
  • Page 157: Reversible Running Belt Rotation For Downhill Walk

    2.51 m (96.5 inches) with 0 % elevation, 2.68 m (105.5 inches) with max. elevation 12.14 Reversible running belt rotation for downhill walk The optional function “reverse running belt rotation” permits simulation of a downhill run.
  • Page 158: Pull Cord Safety Stop Actuator

    Wear appropriate clothing. 12.16 Bottle holder, drinking bottle The bottle holder is available for all h/p/cosmos running machines and can be easily fixed with a clamp on a handrail with 60 mm diameter. The position of the bottle holder should not be at the rear end of the handrail, because this end must be reachable for the subject.
  • Page 159 The note ebook holder ca an be turned to o the field of vis sion of the athlete o r patient.
  • Page 160: Safety Multiple Socket

    The sta able safety mu ltiple socket ca an be used for r the safe connec tion of a medic cal h/p/cosmos treadmill with p peripheral devices s such as PCs, printer or monit tors. It meet s all requireme...
  • Page 161: Rs232 Interface / Usb Converter

    “Technical data“. More recent computers will only have a USB interface instead of the RS232 interface. In that case a “USB to RS232 interface adapter cable“ is available at h/p/cosmos under order number [cos12769]. To control via USB interface, the processor must be a Pentium 1.8 GHz or higher.
  • Page 162 If option 20: 3 is chosen (printer protocol see list of options), data will be printed out as shown in the left example. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 162 of 191...
  • Page 163: Polar Heart Rate Measurement System

    Note: Even when using the coded senders, the transmission of the heart rate to the treadmill is still un- coded. The models h/p/cosmos locomotion, h/p/cosmos pulsar, h/p/cosmos venus and h/p/cosmos saturn are equipped with the POLAR W.I.N.D. transmitter with WearLink technology and coded transmission. The transmitter is only activated if placed correctly on the body.
  • Page 164 Do not clean the electrodes mechanically. Do not use alcohol. If an irregular pulse display occurs in spite of a faultless technical condition, please check your pulse manually or, when in doubt, consult your doctor. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012...
  • Page 165 ® well-known version 3.0. In addition to the pure remote control function of h/p/cosmos treadmill ergometers via the PC, the software now provides additional functions such as an option setting configuration tool with integrated storage function, firmware update over flash eprom, quick-stop key, cool-down key, pause key, target speed settings, target elevation settings, stop countdown, display reset, etc.
  • Page 166: Pc Software H/P/Cosmos Para Graphics

    The treadmill can be controlled automatically. The software is not designed for medical diagnostics or evaluation. The treadmill data can support the doctor’s decision. Data can be converted by the software for data transfer to EXCEL and further evaluattion e.g. by EXCEL or h/p/cosmos para analysis software.
  • Page 167: Pc Software H/P/Cosmos Para Analysis

    The evaluation of heart rate and lactate curves after graded tests is possible on the basis of individually selected threshold value models in graphics and tables. Test data can be entered into h/p/cosmos para analysis manually or by importing from h/p/cosmos para graphics : The entry and ®...
  • Page 168: Pc Software H/P/Cosmos Para Motion

    12.25 PC Software h/p/cosmos para motion ® The h/p/cosmos para motion is a software for recording and evaluating video analysis with individual report for printout, incl. control functions of h/p/cosmos running machines. The software is not designed for medical diagnostics or evaluation.
  • Page 169: Disposal

    13 Disposal Upon request and at the expense of the client, h/p/cosmos might take over the disposal of old or defective devices. Please contact service@h-p-cosmos.com for a detailed offer. Note the information for possible disposal of the running machine parts or components through the client or a subcontractor.
  • Page 170: Appendix 1: Certificates

    14 Appendix 1: Certificates 14.1 Certificate of the TSA (TÜV) according to ISO 9001 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 170 of 191...
  • Page 171: Certificate Of The Tsa (Tüv) According To En Iso 13485

    14.2 Certificate of the TSA (TÜV) according to EN ISO 13485 file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 171 of 191...
  • Page 172: Ec Certificate Of The Tsa (Tüv)

    14.3 EC Certificate of the TSA (TÜV) file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 172 of 191...
  • Page 173: Ec Declaration Of Conformity For Sports And Fitness Devices

    14.4 EC Declaration of conformity for sports and fitness devices file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 173 of 191...
  • Page 174: Ec Declaration Of Conformity For Medical Devices

    14.5 EC Declaration of conformity for medical devices file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 174 of 191...
  • Page 175 © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 175 of 191...
  • Page 176 © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 176 of 191...
  • Page 177 © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 177 of 191...
  • Page 178: Appendix 2

    A serial printer or a PCL printer with parallel/centronics plug und converter [cos10056] can be connected directly to the h/p/cosmos running machine. Following data can be printed after the UKK walk test even without using any PC: UKK fitness index, date, time, test duration, distance, elevation, heart rate, age, sex, body mass index (BMI).
  • Page 179: Instruction Protocol, Checklist

    After the instruction is completed, the instruction protocol is to be signed by the h/p/cosmos technician and all trained persons. Signed instruction protocol, signed delivery note and the registration form are to be sent to h/p/cosmos.
  • Page 180 Operation of running machine via program mode. Refer to the list of programs in the instruction manual. accomplished file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 180 of 191...
  • Page 181 Only connect to coscom v3 - compatible and authorized host equipment. Do not connect to equipment or software which is not explicitly declared as compatible by h/p/cosmos and by the other manufacturer. Inform about the h/p/cosmos software programs h/p/cosmos para control...
  • Page 182 Lubrication of running belt when "OIL"-message is indicated (quantity, intervals according to this manual, distribution of oil by "duck-walking"). Use only original h/p/cosmos lubricant authorized for this model of running machine. Observe additional lubrication requirements. After running belt lubrication, the position of the running belt must be checked once more and be adjusted if necessary.
  • Page 183 Advice and support documents Hand over and explain delivery papers (folder). Fill out registration form and send it to h/p/cosmos sales. Let the customer sign the delivery note and instruction protocol and send it back to h/p/cosmos sales immediately.
  • Page 184: Instruction Protocol, Signatures

    This will furthermore result in loss of liability and warranty. Please fill out the instruction protocol and send it back to h/p/cosmos via Fax to +49 86 69 86 42 49, via email to sales@h-p- cosmos.com or via post...
  • Page 185: Service Report – [Cos15531]

    15.4 Service report – [cos15531] file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 185 of 191...
  • Page 186: Control Protocol – [Cos11690En-02]

    15.5 Control protocol – [cos11690en-02] file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 186 of 191...
  • Page 187: Wiring Diagram Load Circuit Quasar ® Med 5.0 / Pulsar ® 5.0

    15.6 Wiring diagram load circuit quasar med 5.0 / pulsar ® ® Color Code neither US / nor Canadian standard! file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 187 of 191...
  • Page 188: Wiring Diagram Control Circuit Quasar ® Med 5.0 / Pulsar ® 5.0 – Page 1

    Wiring diagram control circuit quasar med 5.0 / pulsar 5.0 – page 1 ® ® Color Code neither US / nor Canadian standard! file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 188 of 191...
  • Page 189: Wiring Diagram Control Circuit Quasar ® Med 5.0 / Pulsar ® 5.0 – Page 2

    Wiring diagram control circuit quasar med 5.0 / pulsar 5.0 – page 2 ® ® Color Code neither US / nor Canadian standard! file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 189 of 191...
  • Page 190: Appendix 3: Symbols

    Warning of hot surfaces electronic equipment. Necessary to reduce waste management problems linked to heavy metals and Risk/danger of entanglement flame retardants. file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 190 of 191...
  • Page 191: Contact

    & production Am Sportplatz 8 DE 83365 Nussdorf-Traunstein Building 2 (picture below) h/p/cosmos sales & service Feldschneiderweg 5 DE 83365 Nussdorf-Traunstein file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx © 2012 h/p/cosmos sports & medical gmbh author: m email@h-p-cosmos.com created 04.05.2012 printed 04.05.2012 page: 191 of 191...

Table of Contents