OP12
Setting of allowance for speed signal
comparison (measurement / control)
These values do not reflect the real
speed accuracy of the running belt.
Accuracy of running belt speed is
higher.
OP13
Activating and deactivating motor
brake
OP14
Maximum acceleration time for
RS232
(option only available for treadmills,
not for ladder ergometers)
OP16
Activating and deactivating heart
rate measurement
© 2012 h/p/cosmos sports & medical gmbh
The MCU5 control unit continuously compares the data of the speed
measurement with a calculated reference value. In order to detect
problems with the speed measurement or speed control or e.g.
performance limitations due to a bad voltage supply, an allowance
range can be set in this option. If the allowance is exceeded, the
error message E.31 (measured speed is too high) or E.32 (measured
speed is too low) is indicated
§
6 = 6 % deviation is allowed
8 = 8 % deviation is allowed
10 = 10 % deviation is allowed
The motor brake prevents the running belt from moving manually (for
example due to the subject's bodyweight) when the speed value is set
to "0". If the treadmill recognizes running belt movement, it activates
the motor break for 10 seconds.
Remark:
The motor brake works only on running machines with integrated
speed measurement. The speed sensor must be switched on
(OP11=1). If the running machine is not equipped with a speed
measuring system, this can be retrofitted easily.
To activate the motor brake, some parameters of the frequency
inverter must be changed. Ask the service department of h/p/cosmos
sports & medical gmbh for detailed instructions.
§
§
§
$
%
=
0 = heart rate measurement is deactivated / OFF
1 = heart rate measurement is activated, circuit board HFU1 or
RE06REC (default)
2 = heart rate measurement is activated, circuit board HFUi
auto = automatic detection and activation of circuit board
4 = POLAR W.I.N.D. (further settings in user OP19 and OP23
necessary)
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
optional settings: administrator options
indicates: OFF = no comparison
Attention!
If the device is switched off or disconnected from
the power supply, the motor brake will not
function!
After 10 seconds of operation of the motor brake,
the brake is automatically released and will only be
activated again after belt movement. A short jerk
may occur during reactivation.
shows OFF = motor brake not active
shows ON = motor brake active
shows e.g. 5 (max. acceleration increment)
shows e.g. 131.1 (max. acceleration time in sec.)
shows: SEC.
reports ...
created 04.05.2012
printed 04.05.2012
page: 125 of 191