Page 2
These instructions for use are only valid for the original configuration of the first delivery of the devices pictured below. If the device has been changed, please refer to the latest version of this document, available at: www.h-p-cosmos.com Product family: treadmill h/p/cosmos 150/50 LC Models: pluto pluto lt med ®...
Page 3
These instructions for use include all of the information needed to operate the device correctly and safely. Please read them carefully before use and keep them available at all times. We hope you will have a lot of fun and success as you work with your h/p/cosmos device. Franz Harrer President & CEO h/p/cosmos sports &...
1 Symbols and Labels 1.1 Symbols used (general) Illustraion Description Reference CE sign, proof that the essential requirements (according to medical device directive (here with number of notified body) were met 93/42/EEC) General warning (danger, warning or caution statements) (acc. to DIN EN ISO 7010 W001) Warning of obstacles (stumbling) (acc.
1.2 Symbols used (transport) Illustraion Description Reference Fragile, Handle with care (acc. to ISO7000-0621) This way up (acc. to ISO7000-0623) Keep dry (acc. to ISO7000-0626) Centre of gravity (acc. to ISO7000-0627) Temperature limitations (acc. to ISO7000-0632) Do not stack (acc. to ISO7000-2402) page 8 cos105000-2016-00 en...
Gait training (with or without body weight support) ❚ h/p/cosmos medical treadmills can be used in combination with external devices for walking or running in place as Loading devices for neuromuscular and biomechanical measurements (e.g. EEG, EMG, motion analysis) ❚...
4 Safety h/p/cosmos medical treadmills may be operated with healthy subjects as well. For applications with healthy subjects, please apply the instructions for use for sports devices, available at www.h-p-cosmos.com 4.1 Safety information – Forbidden use Obey the following danger, warning and caution statements stricktly in order to prevent serious injury or death! For any medical treadmill application, a fall prevention device is prescribed and absolutely obligatory.
Page 14
In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection, ❚ mark clearly and inform h/p/cosmos service personnel via telephone and writing. In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service ❚ personnel via telephone and writing.
For any medical treadmill application, a fall prevention device is prescribed and absolutely obligatory. h/p/cosmos provides fall prevention devices in the form of a safety arch or a body weight support device (airwalk). It is up to the operator to use any other device that prevents the patient from falling and complies with IEC60601-1.
Page 16
Description Ilustration Tighten shoulder and chest straps. Use carabiner to connect safety harness with rope. Explain function to subject. Adjust the length of the rope in a way the patient has to maintain the correct position (see “position of patient and operator”).
4.3 Emergency dismount Patient is conscious and aware of danger. Patient grabs the handrails ❚ Patient jumps off the running belt onto the foot rails ❚ Operator / patient hits the emergency off ❚ Patient is conscious but not aware of danger. Patient stumbles and falls into fall prevention device.
4.4 Emergency off Do only use in case of emergency. Do not use as normal stop button. Do not stand on or enter the running deck when device is in elevation (running belt might slip through due to gravity). Operation Result Release Restart...
4.6 Unauthorized access See OP 40 … 44 in the User Options to lock the whole device or individual modes. 4.7 Residual risk / Side effects After risk reduction only 13 of more than 230 causes are in the "widely acceptable" region. In case fall prevention is not applied or not applied correctly, there are residual risks, such as falling of a person resulting in skin abrasions, bruises, fractures or in worst case even death.
5 Preparation Description Ilustration Perform daily inspection as described in “daily inspection”. Explain device and application to patient. Explain and apply fall prevention as described in “fall prevention”. Guide patient onto treadmill. Do not enter the device when running belt is rotating. ❚...
Page 21
Description Ilustration Apply pull-cord safety stop. (Attach clip to patients’ clothing.) Adjust the length of the rope so that the patient has to maintain the correct position (see “position of patient and operator”). page 21 cos105000-2016-00 en...
6 UserTerminal 6.1 Keys and displays Primary function Element Secondary functions Displays currently activated mode Displays current speed in m/min, km/h, m/s or mph (see LED)* Displays max. speed when selecting modes Displays covered distance in m, km or miles (see LED)* Displays elapsed time in minmin:ss or hh:minmin Displays duration when selecting profiles Displays current profile step / number (see...
6.2 Standard vs. “lt” devices Most h/p/cosmos treadmills are available as standard or “lt” (light) devices. “lt” devices have no UserTerminal (no display, no keyboard). Standard device “lt” device Control via Control via - UserTerminal - Software (see “accessories”) - Software (see “accessories”) - additional keyboard (see “accessories”)
6.3 Connection of external devices / Interfaces The back of the UserTerminal has RS232 interfaces. For USB-RS232 converter, see “accessories”. The person combining a medical device with any other device for the first time is creating a Medical Electrical System. Requirements for ME-Systems, see “creating systems”.
7 Position of Patient and Operator Position of patient (initial contact) 1a) Optimal position 40%, front 1b) Tolerated position 30%, middle running area 1c) Not tolerated position / buffer zone 30%, rear Intended position of operator The operator must be within reach of the emergency off at all times. If the operator is not able to reach the emergency off button at the UserTerminal (body height, obstacles, etc.), the operator must install an additional emergency stop within reach (see accessories).
8 Operation 8.1 General application procedure Description Illustration Disinfect the device (see “cleaning”). Disconnect the device and all accessories from mains power supply before cleaning or disinfection. Make sure… …the PE-cable is connected to electrical installation and device, …the device is directly plugged into the dedicated wall socket, …the device protection switch on the front of the device is switched on (light on),...
8.3 Manual mode Basic functions Buttons / displays Further information Device must be in “mode selection” Select “manual mode” with “+” or (one of the mode LEDs flashes) “-“ button To get there, cancel all other Selected mode flashes. activities by pressing the “stop”- button.
8.4 Profile mode The automatic modes must only be performed on the prescription of the medical doctor. ❚ Operator and patient have to be aware of automatic load changes during profile, cardio and test mode. ❚ Start and load changes within the automatic modes are indicated by acoustic signals (beep). Furthermore, the displays show the next load parameters (flashing).
Page 30
Basic functions Buttons / displays Further information Device must be in “mode selection” Select “profile mode” with “+” or “-“ (one of the mode LEDs flashes) button. To get there, cancel all other Selected mode flashes. activities by pressing the “stop”- button.
In case of any visible or assumed defects or malfunctions (of the device, accessories, software, etc.), unplug device, exclude ❚ reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing. Start and load changes within the automatic modes are indicated by acoustic signals (beeps).
Page 32
If the heart rate signal totally fails, an acoustic warning signal occurs and the heart rate display shows no value any more. Furthermore, the device reduces speed and elevation to 0 within one minute. For possibilities to interfere with an automatic program, see “interfere with an automatic program”. The treadmill adjusts speed and elevation according to following matrices.
In case of any visible or assumed defects or malfunctions (of the device, accessories, software, etc.), unplug device, exclude ❚ reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing. Start and load changes within the automatic modes are indicated by acoustic signals (beeps).
Page 34
Basic functions Buttons / displays Further information Following tests require further settings: Select test with “+” or “-“ 01 UKK walk test Confirm with “enter”. 02 Graded test The running belt starts automatically after countdown. 03 Conconi test 10 Ramp profile Terminate the operation with “stop”.
8.7 Interfere with automatic program Basic functions Buttons / displays Further information Profile / test mode: Only valid for current step. Alter speed Cardio mode: Press “+” or “-“. Reduce the speed with “-“ or exceed the max. speed with “+”; This sets a new max.
8.8 Pause function Reducing the speed with the “-“ key to 0.00 km/h, triggers the “pause” function. The running belt stops. The speed display shows “PAUS”. Pressing the “start” key, starts the running belt at the preset starting speed (default: 0.5 km/h). All values are maintained. Pressing the “+”...
8.9 Acceleration levels Start the use of the treadmill with slow walking, especially for beginners. ❚ There are seven acceleration / deceleration levels for any kind of operation. The acceleration levels are defined by the time it takes to accelerate from 0 km/h to maximum speed. Example: With acceleration level 3, it takes 33 seconds from 0 km/h to maximum speed (see table below).
8.10 User Options Basic functions Buttons / displays Further information Press “+”, “-“ and “stop” together for “speed “ display shows “OP 01”. at least 3 s. Select option with “+” or “-“ Confirm with “enter”. Adjust option with “+”or “-”. Confirm with “enter”.
Page 39
RS 232 interface protocol: COM 2 OFF, 1 ... 18 (1 = h/p/cosmos coscom) OP 23 RS 232 interface protocol: COM 4 OFF, 18 … 23 (20 = h/p/cosmos coscom v3) OP 27 Min. acceleration and deceleration level 1 ... 5 OP 28 Max.
It is obligatory to perform a risk management when creating an ME-System / PEMS. Risk management, safety, compliance, and maintenance are the responsibility of the manufacturer of the ME-system / PEMS, not the responsibility of h/p/cosmos. For all ME-Systems / PEMS a potential equalization must be provided.
(*) At least one kind of fall prevention is mandatory, see “safety”. 9.3 Compatible devices A number of ECG and CPET (cardiopulmonary exercise test) devices as well as software products are compatible with h/p/cosmos treadmills via coscom v3 interface protocol.
Clean the device regularly − Unplug the device. − Use a damp cloth (not wet). − Wipe all surfaces. − Wash safety harness acc. to label. The h/p/cosmos devices are neither sterile nor can they be sterilized. page 42 cos105000-2016-00 en...
In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection, ❚ mark clearly and inform h/p/cosmos service personnel via telephone and writing. In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service ❚ personnel via telephone and writing.
For further and precise analysis a detailed instructions regarding the tissue lubrication test are available at service@h-p-cosmos.com. Fill the syringe with 10 ml of the lubricant (h/p/cosmos special oil, only) and attach the tube to the syringe. Pump the lubricant under the running belt very slowly and remove the syringe.
Description Illustration Start the running belt with 2 km/h and walk on the belt for 2 min. Vary the position in order to distribute the lubricant. Reset the OIL message (see OP01 in “User options”). 11.4 Adjustment of running belt Unmeant trapping hazards: Take off ties, scarfs or other clothes that may be trapped.
All maintenance work that is not explained in detail, must not be performed by the operator. Safety inspections, installation and repair work, must also not be performed by the operator. This kind of work must be performed by h/p/cosmos service personnel according to the “h/p/cosmos service instructions”. The “h/p/cosmos service instructions” are available at service@h-p-cosmos.com.
In case of any visible or assumed defects or malfunctions (of the device, accessories, software, etc.), unplug device, exclude ❚ reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing. In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection, ❚...
12.2 RS232 troubleshooting Problem Solution No connection via RS232 For connection to PC with h/p/cosmos software and most external devices use the included RS 232 interface connection cable [cos00097010034]. (wrong cable) No connection via RS232 Check cable and plugs for defects.
13 Technical data 13.1 UserTerminal Description Data Displays 6 seven-segment LCD displays, LED displays of mode and unit Keyboard 6-key keyboard film “lt” devices have no UserTerminal (no display, no keyboard). 13.2 Dimensions Description Data Device dimensions 2095 x 850 x 1310 mm (L x W x H) Running surface dimensions 1500 x 500 mm (L x W) Track access height...
13.6 Environmental conditions Operation Data Temperature 0 … 40° C Humidity 20 … 90%, without condensation Pressure 700 … 1060 hPa Altitude max. 3000 m, without pressurization (altitudes >1000m can cause minor loss of performance) Oxygen saturation <= 25% Exclude danger of explosions or inflammability. Exclude high voltage lines / devices in near vicinity. Transport &...
13.9 Classification Description Data MDD 93/42/EEC Notified body MDD 93/42/EEC Risk class active therapeutic device and active diagnostic device 60601-1 Protection against electric Class I, shock IEC 60601-1 Protection against harmful IP00 ingress of water or particulate matter IEC 60601-1 Mode of operation Continuous operation IEC 60601-1...
13.12 Voltage, Current, Performance Description Data Input voltage * 200 V - 240 V ~ (f: 50 – 60 Hz) Current input (long time) * 6.0 A Current input (momentary) * 13.5 A Energy consumption (long time) ≤ 1440 VA Energy consumption (momentary) ≤...
Limited liabilities apply: If h/p/cosmos or h/p/cosmos organizational bodies, senior management or agents can be held accountable for the payment of damages pertaining to slight negligence (breach of material contractual obligations), the damages shall be limited to damages that could typically have been foreseen. Liability pertaining to slight negligence excludes liability as a result of loss of production, interruption of business and loss of profits.
17.2 Instruction protocol When installing the device, the h/p/cosmos service personnel instructs the intended operator according to the instructions for use, following this instruction protocol. With the name and signature on the instruction protocol, the instructed persons confirm that they fully know how to operate the device safely.
Furthermore, this will result in loss of liability and warranty. Please fill out the instruction protocol and send it back to h/p/cosmos by fax (+49 8669 8642 49), email (sales@h-p-cosmos.com) or mail).
Comment / Display OP01 Reset (delete) error messages The required maintenance work must be performed before deleting error messages. The h/p/cosmos service department must be contacted before one of the service displays or error codes is deleted! Confirmation on display indicating "donE"...
Page 57
Option Description Comment / Display OP11 Scaling of the profiles in profile mode (not for shows: Scaling possibilities test mode) 0: No scaling (standard) 1: Scaling 1...6, which is shown in the profile mode on the INDEX display, refers to all parameters together (speed, elevation, time) 2: Scaling 1...6 refers to each parameter individually (speed, elevation, time) OP12...
Page 58
OFF = RS232 not active / no protocol / interface deactivated h/p/cosmos coscom v1, v2, v3 with baud rate 9600 bps (standard setting COM 1 and COM 2) printer protocol (serial printer or converter required) TM treadmill emulation in km/h...
Page 59
28. Note: If the peripheral equipment sends an acceleration and deceleration command via the h/p/cosmos coscom protocol, the selected level in option 29 or in the corresponding command of the h/p/cosmos coscom protocol.
Page 60
Option Description Comment / Display OP48 Program step countdown OFF = The time display counts up each program step ON = The time display counts down each program step OP52 Output interval for printer protocol By entering a value between 0 and 100, the output interval is set in seconds for a printer directly connected to the treadmill.
18 Annex II (Pre- & self-defined tests) 18.1 UKK walk test UKK stands for Urho Kaleka Kekkonen, founder of the UKK Institute in Tampere, Finland. The UKK walk test is a fitness test, calculating the UKK Fitness Index based on the measured heart rate within a 2 km walk at max. walking speed.
18.2 Graded test Description Illustration (e.g. for performance diagnostics based on lactate measurement) Refer to “test mode” for safety information and adjustment. Parameter Default value Starting speed 8.0 km/h Increment 2.0 km/h Acceleration level Step length 3:00 min Break time 0:30 min Each parameter is adjustable.
18.4 Bruce protocol Description Illustration e.g. for ECG stress test Refer to “test mode” for safety information and adjustment. Step Duration (min) Speed (km/h) Elevation (%) 3:00 18.5 Naughton protocol Description Illustration e.g. for ECG stress test Regard chapter “test mode” for safety information and adjustment. Step Duration (min) Speed (km/h)
18.7 Cooper protocol Description Illustration e.g. for ECG stress test Refer to “test mode” for safety information and adjustment. Start at 5.3 km/h and 0% elevation ❚ After 1 minute elevation increases to 2 % ❚ After another minute, the elevation is increased by 1% every minute ❚...
18.10 Ramp profile Description Illustration (not available for every model) Refer to “test mode” for safety information and adjustment. Ramp profile with 2 parameters: Target speed standard: 10.0 km/h; adjustable from 0 to maximum ❚ speed of the treadmill. Time for reaching target speed in seconds: standard: 10 seconds; ❚...
18.12 Self-defined tests Tests 21 – 28 are freely definable with up to 40 program steps. Follow the instructions below to program an individual test. The “heart rate” display will show the current program step. Use “up” and “down” to scroll through the program steps. Basic functions Buttons / displays Further information...
Page 67
Basic functions Buttons / displays Further information Select acceleration for step 1 The acceleration level is displayed with “+” or “-“. in the “index” display. Confirm with “enter”. Proceed the same way for further steps. Make sure all steps after the last one have a speed of “0 km/h”.
19.1 Arm support, adjustable [cos12013] Title Description Short description The h/p/cosmos arm supports are a simple solution for unweighting of the patient. Height and width adjustability offers a wide field of application. Illustration Application Adjust the arm support by pulling the locking element and turning the segments.
UserTerminal Push button Release button displays “pull stop” Restart application Mains connection and interface communication not interrupted Additional safety information Technical data Additional accessories Installation By h/p/cosmos service personnel, only Further information https://www.h-p-cosmos.com/en/products/individual-products/additional-stop-button-right page 69 cos105000-2016-00 en...
Patient may hold crossbar front rail for balance control Holding handrails during use affects exercise results. Additional safety information Technical data Length: 700 mm Diameter: 40 mm Weight: 1.3 kg Technical data Installation By h/p/cosmos service personnel, only Further information https://www.h-p-cosmos.com/en/products/individual-products/handrail-crossbar-5036- arched-grip-cover page 70 cos105000-2016-00 en...
19.4 Elevation 0% to +25% [cos102927] Title Description Short description Extends elevation to 25% (22.5°) Illustration Application Additional safety information Technical data Max. elevation: 22.5° Additional accessories Installation No retro-fitting Further information page 71 cos105000-2016-00 en...
Patient may hold handrails for balance control. Holding handrails during use affects exercise results. Additional safety information Technical data Length: 1358 mm Diameter: 40 mm Weight: 9.5 kg (4.0 kg additional weight) Additional accessories Installation By h/p/cosmos service personnel, only Further information page 73 cos105000-2016-00 en...
Length: 910 mm Width: 855 mm Height: 543 mm Max. patient weight: 50 kg Max patient weight of treadmill is reduced when combined with arm support. Additional accessories Installation By h/p/cosmos service personnel, only Further information page 74 cos105000-2016-00 en...
POLAR chest belt XS cos10906 POLAR chest belt S cos10165 POLAR chest belt M cos10907 POLAR chest belt L cos10902 POLAR transmitter set T31 cos15178 POLAR transmitter set T34 (extended range) Installation By h/p/cosmos service personnel, only Further information page 75 cos105000-2016-00 en...
- Other chest belt than POLAR W.I.N.D. is used (see print) - Device and chest belt are not connected (see User Option OP 19) Additional accessories cos100420b POLAR WIND transmitter TRX24 cos100420c POLAR WIND WearLink chest belt Installation By h/p/cosmos service personnel, only Further information https://www.h-p-cosmos.com/en/products/individual-products/polar-heart-rate-receiver-wind page 76 cos105000-2016-00 en...
USB-RS232 converter cos00097010034 Interface connection cable RS 232 5 m cos00097010035 Interface connection cable RS 232 10 m Installation By operator Install from h/p/cosmos demo & info DVD or download from website (see below). Further information https://www.h-p-cosmos.com/en/products/software/hpcosmos-para-control-410 page 77 cos105000-2016-00 en...
19.11 robowalk expander [cos30022, cos30023] Title Description Short description The h/p/cosmos robowalk expander supports gait training. Expander ropes, attached to the limbs, support or load the patient. Illustration Application See separate instructions for use. Additional safety information See separate instructions for use.
19.12 Safety arch for treadmill families 150/50 LC and 150/50 [cos10079] Title Description Short description The h/p/cosmos safety arch one option to secure the patient from falling. Furthermore, the safety arch stops the treadmill in case of falling. Illustration Application See “fall prevention”...
19.13 Wheelchair ramp [cos102931] Title Description Short description Wheelchair ramp supports entering the device with wheelchair patients. Illustration Application Push patient with wheelchair onto treadmill. Connect patient to fall prevention device. Support patient so they can stand upright. Remove wheelchair. Start application.