Electrical Installation - h/p/cosmos 150/50 mercury Operating And Service Manual

Running machine series
Table of Contents

Advertisement

7.4.6

Electrical installation

An overload or voltage drop (even temporary) of more than 20 % of the mains voltage might cause
a malfunction and/or defect and might totally switch off the running machine. In the event of such
high voltage drops or total power failure, the running machine switches off and the running belt
stops. To start again, switch on the running machine at the main "ON" switch and press START at
the keyboard.
h/p/cosmos running machines are to be installed at a voltage power connection with ground wire (earthing,)
N
including leakage-current protection switch (interrupter) and according to VDE 0100 or/and the currently valid
regulations and directives. Requirements for special locations, areas and establishments (e.g. medically used
areas) must be strictly adhered to. PE connection (protection-earth contact) is stipulated for all running machines.
The voltage drop between the beginning of the consumer's installation and the wall socket must not be more than
4 % (DIN VDE 0100-520). It is the consignee's and user's personal responsibility to check the correct function of
the mains connection including the outlets. An authorized electrical engineer must check these points for perfect
functioning regularly (1 ... 4 years). Inspections of the electrical installations within the building are not incumbent
on the supplier h/p/cosmos.
Read the nameplate of the machine before connecting the machine to the power outlet in order to determine
N
whether the machine was designed for special voltage supply (e.g. 110 Volts and 25 Ampere). The standard
power supply of 230 Volt / AC, 50/60 Hz is sufficient for most of the running machines.
The bigger running machines (with a size of running surface 190/65 cm models
N
pulsar 3p and venus / saturn up to 300/125 cm) require a 3 phase AC power supply:
3 x 400 Volt/32 Amp fuse (16 Amp for models pulsar 3p) and clockwise rotation.
right: CEE power outlet (socket), 5-pin, 16 Ampere
[cos11092]
left: CEE power outlet (socket), 5-pin, 32 Ampere
[cos11090]
Power supply for h/p/cosmos pulsar
3p
400 Volt AC 3~/N/PE f: 50/60 Hz Imax:
16.0 A
3 phase current with clockwise rotation
field.
16 Ampere fuse (C characteristic)
The mentioned wire colours are applicable for Germany; different colours may be applicable for other countries.
Voltage values for 3-phase voltage power supply:
from L1 to L2: 400 Volt, from L1 to L3: 400 Volt, from L2 to L3: 400 Volt
from L1 to N: 230 Volt, from L2 to N: 230 Volt, from L3 to N: 230 Volt,
© 2012 h/p/cosmos sports & medical gmbh
Power supply for h/p/cosmos venus
and saturn
400 Volt AC 3~/N/PE f: 50/60 Hz Imax:
32.0 A
3 phase current with clockwise rotation
field.
32 Ampere fuse (C characteristic)
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
safety, installation and commissioning
created 04.05.2012
printed 04.05.2012
page: 81 of 191