12.20 RS232 interface / USB converter

Connecting a running machine to a medical device results in a medical system.
Always use IEC 60601-1-approved potential isolation components when linking medical devices
via interface.
Only authorised trained staff is authorized to perform this connection.
This medical system has to be connected via a potential balance equalization cable with the
provided connector bolt and bearing in the designated room.
Standard for all devices is ONE serial interface COM1. Optionally further serial interfaces COM2, COM3 and COM4 are
available. For further information, please contact your local dealer or h/p/cosmos. The interface protocol h/p/cosmos
coscom
is available on request or can be found at www.coscom.org. The incorporated protocols (e.g. for ECG-devices)
©
are listed under OPTION 20 / OPTION 21. See the compatibility list of the RS232 interface in the chapter entitled
"Technical data". More recent computers will only have a USB interface instead of the RS232 interface. In that case a
"USB to RS232 interface adapter cable" is available at h/p/cosmos under order number [cos12769]. To control via USB
interface, the processor must be a Pentium 1.8 GHz or higher.
Interface cable RS232 for connection to PC
for h/p/cosmos para graphics
1
6
2
7
3
8
4
9
5
PC COM1 oder COM2
CONNECTOR DB9
Buchse / female
[cos00097010034] 5 m cable, [cos00097010035] 10 m cable
Interface cable RS232 for connection to ECG Schiller (Schiller AT 10 / AT 60 or CS 100 / CS 200)
1
6
2
7
3
8
4
9
5
EKG / ECG Schiller
CONNECTOR DB9
Stecker / male
[cos00097010036] 5 m cable
Interface cable for connection ergospirometry JAEGER / VIASYS / CardinalHealth / CareFusion OXYCON
1
6
2
7
3
8
4
9
5
OXYCON
CONNECTOR DB9
Buchse / female
[cos00097010034] 5 m cable, [cos00097010035] 10 m cable
Connect the running machine interface with an available serial interface (COM1 ... COM4) at the PC of the
OXYCON. Do not use the female connector at the OXYCON with the letters "treadmill/running machine".
Important: As of January 2000 use h/p/cosmos coscom
, h/p/cosmos para control
®
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
© 2012 h/p/cosmos sports & medical gmbh
, PC, ECGs, ergospirometry devices
®
1
R x D
2
T x D
3
4
GND
5
Laufband / treadmill
CONNECTOR DB9
Stecker / male
1
R x D
2
T x D
3
4
GND
5
Laufband / treadmill
CONNECTOR DB9
Stecker / male
1
R x D
2
T x D
3
4
GND
5
Laufband / treadmill
CONNECTOR DB9
Stecker / male
RS232 interface protocol.
©
email@h-p-cosmos.com
created 04.05.2012
accessories and options
9-pole Sub- D
6
(male - female)
7
with crossed
8
9
transmission-
receive-cable
(PIN 2 and PIN 3)
6
7
8
9
OXYCON
Alpha / Delta /
6
Champion / Pro
7
8
9
printed 04.05.2012