When connecting treadmills to the multiple safety socket, there are no performance limitations to expect at low-
N
performance applications (walking, slow running). For high-performance applications (fast running, sprints, etc.,)
the use of the multiple safety socket can result in performance limitations.
For treadmills used in the medical field:
N
All devices are to be connected via a potential compensation cable in „star connection"with the potential
N
compensation bar.
First connect the potential compensation (potential equalization) with the corresponding plug pin (next to the
N
main switch on the front) and then connect the mains plug.
During electric safety measurements and tests (leakage currents, etc.) the potential equalization cable has to
N
be disconnected temporarily.
Due to the very high electrical load of treadmills, do not connect more than one treadmill or
N
device with high load to the multiple safety socket. Use a dedicated line only.
After adjusting the running belt, the allen key must be pulled out of the screw immediately after usage due to risk of
N
injury!
There are dangerous capture areas/gaps at the elevation system as well as at the rear end and the sides of the
N
running belt. At reverse belt rotation there is also danger at the motor hood and front area.
It should be ensured that all cables (power connection, interface, potential equalization, etc.) and accessories are
N
installed properly and safely and that nobody can stumble or fall over the cables and/or accessories.
The connection of the gas pressure container has to be checked for MOT-admissibility (TÜV).
N
Absorption, cleaning and disinfecting chemicals are to be kept in the appropriate reservoirs.
N
All lubrication material and all other parts should be kept away from children and animals.
N
Do not modify the running machines and do not connect to other equipment which is not explicitly declared
N
compatible by all involved manufactuers.
Only connect accessories, software and host equipment if it is confirmed as compatible by all manufacturers.
N
It is not allowed to connect sports treadmills to medical devices such as ECG-systems.
N
Connecting a running machine to a medical device results in a medical system. Only trained staff is authorized to
N
perform this connection. Always use IEC 601-1 approved potential isolation components. This medical system is to
be connected via a potential balance cable with the provided connector bolt and bearing within the designated
room.
The power consumption of the connected devices must be in total lower than 300 W.
N
7.1.3

Maintenance and safety inspections

Maintenance and repair of the devices (also opening of the device) have to be performed by service engineers
N
authorized and certified by h/p/cosmos, preferably within the scope of a maintenance contract.
In case of any detected and/or assumed malfunction and/or defects and/or unreadable safety labels, the device is
N
to be disengaged immediately. The device is to be marked and secured against operation and the supplier and
authorized service personnel has to be informed in writing immediately.
Before intervenion in the device, switch the running machine off and pull the mains plug from the power supply.
N
During all maintenance work and safety tests, ensurethat no third parties are directly or indirectly in contact with
N
the device under test and/or the technician performing the test. Keep a safety zone of 2 m radius clear.
Due to environmental influences, interior climate and type of application (for example frequent short-run operation
N
or infrequent long-run operation, etc.), the required automatic oil setting of the treadmill can deviate extensively
from the factory setting.
In case the required lubrication, quantity and quality is not assured due to lack of inspections by the user and/or
N
lack of maintenance by authorized h/p/cosmos service technicians, early damage to the running belt and running
surface can occur and warranty will expire.
© 2012 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
safety, installation and commissioning
created 04.05.2012
printed 04.05.2012
other
electrical
page: 73 of 191