8.10

Hygiene and cleansing

For safety reasons, switch off the device and pull the plug before cleansing, disinfecting and
N
before opening.
Pay attention to the safety notes, warnings and precautions of the running machine, the
N
accessories and the disinfectant.
The h/p/cosmos devices are neither sterile nor can they be sterilized.
Pay attention to the operating instructions of the respective devices and accessories. Chemicals needed for application
or cleansing should be stored in appropriate reservoirs so as not to confuse them.
The surfaces of the treadmill can be cleaned with a light moistened cloth. For disinfection h/p/cosmos recommends
Bacillol plus which can be ordred from h/p/cosmos, order number [cos12179]. Before use of cleaning and disinfection
substances, always test the compatibility of the substance at poorly visible places.
In the event that patients with communicable diseases use the treadmill, the treadmill must be disinfected at the
discretion of the doctor or medical stuff before and after the treatment. This must also be done when a communicable
disease is merely presumed. This covers all parts the patient may have contacted, such as front and side handrails, arm
support, UserTerminal (keyboard and display), emergency stop, chestbelt or other heart rate measurement systems,
running belt, emergency stop with magnet switch, safety harness for fall stop prevention, vest of unweighting system,
motor cover, frameparts, treads, wheelchair ramp, drinking bottle, etc.
8.10.1
Cleansing of outside and applied parts
Spray contact surfaces, impact, wipe and rub off. Spray disinfectant onto a tissue and rub clean any surfaces that may
be contaminated.
Always follow the manual of the disinfectant manufacturer, especially the safety warning and regulations regarding
disposal.
8.10.2
Cleaning of the interior
The interior should be cleaned every 6 months. The fans inside the machineand the moving running belt accumulate
dust and sweat. For this reason frequent interior cleaning is highly recommended for all treadmill models.
Depending on your model, lift up the motor hood or remove bellow from upper frame. Clean the interior of the running
machine by removing dirt with the help of a vacuum cleaner. Pay special attention to the ventilation net of the drive
motor and the cover of the frequency inverter. For devices with an external control unit, the internal space of the control
unit must be cleaned as well.
8.11
Cleansing and adjustment of the speed sensor / light barrier
Depending on the model, the firmware version and the manufacturing date, the machines may be equipped with...
no speed sensor (thus no cleansing required)
N
light barrier and incremental waver mounted without housing at the front (drive shaft) of the drive engine.
N
devices manufactured since appr. 2003, have an induction sensor with metal toothed lock washer instead of light
N
barriers
only for oversize running machines with a running surface of 200/75 cm up to 300/125 cm and built before 2002:
N
the speed sensor is mounted in its own housing at the rear of the drive engine (drive shaft), thus it must not be
cleaned.
The light barrier of the drive motor is a pulse generator for the control unit. The light barrier and the incremental disc
should be cleaned carefully with a cloth moistened with alcohol at intervals of 6 - 12 months. If the disc is housed, you
© 2012 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
maintenance and safety inspections
created 04.05.2012
printed 04.05.2012
page: 110 of 191