Devices Without Userterminal And Without Automatic Oil Pump - h/p/cosmos 150/50 mercury Operating And Service Manual

Running machine series
Table of Contents

Advertisement

[03]
[04]
After all required steps have been performed, return the tissues to h/p/cosmos for evaluation. The tissues
will be weighed again. A weight difference indicates whether the lubrication interval needs to be adjusted.
8.4.3

Devices without UserTerminal and without automatic oil pump

Devices without a UserTerminal are equipped with an automatic acoustic signal (bleep) indicating the device
needs to be lubricated. At the standard setting, the acoustic signal will occur after every 1000 km of operation.
(Special devices may have different intervals.) After switching on the device using the main power switch, the
acoustic signal will be repeated three times: 5 times LONG (code for "0"), 1 time SHORT and 4 times LONG (code
for "1"). The device has no sensor; the oil message does not disappear automatically after lubrication.
Lubricate the running belt and running surface with the enclosed bottle of lubricant and 10 ml syringe, included in
the scope of delivery. Do not use other lubricants as they may damage the running belt and surface and cause a
breakdown of the running machine.
Please contact h/p/cosmos to order the special oil. The running belt should be lubricated after 1000 km at the
latest or/and if dry grinding noises are detected during operation. Special devices may have different intervals.
After lubrication the oil message should be reset with "OP 01". This requires an external UserTerminal or PC
software h/p/cosmos para control
8.4.4
Devices with UserTerminal & automatic oil pump (models "r" for cycle & wheelchair)
Running machines which are designed for cycling and wheelchair use are equipped with an automatically
controlled oil-pump and an oil tank.
The oil tank is located in the external control unit of the device below at the left. For devices constructed before
2002 and without external control unit, the oil tank is located at the rear of the machine, close to the drive engine.
© 2012 h/p/cosmos sports & medical gmbh
.
®
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
maintenance and safety inspections
1. Take the tissue out of the plastic bag and
place it directly on the previously marked
position under the running belt.
2. Press your hand on the running belt above the
tissue and move the running belt together with
the tissue for exactly 1 meter.
3. Remove the tissue and put it into the
corresponding plastic bag. Take care that the
plastic bag is closed properly.
Repeat steps 1 to 3 for the remaining positions.
created 04.05.2012
printed 04.05.2012