h/p/cosmos 150/50 mercury Operating And Service Manual page 95

Running machine series
Table of Contents

Advertisement

[05]
[06]
[07]
[08]
[09]
© 2012 h/p/cosmos sports & medical gmbh
Perform an electric safety test/measurement for
N
N
N
Record the test results in a special protocol [cos11690xx] and
mark them as "first measured values". One copy of this protocol
remains with the owner's manual and the original is sent to
h/p/cosmos.
The potential equalization cable has to be connected AFTER these
measurements.
Lubricate the running machine with 30 ml special silicone oil, which
is part of the "service box".
Check the tension of the running belt (no slip).
Adjust the running belt properly before instructing the customer.
Test the function of the heart rate transmission with the help of a
simulator or a POLAR sender. Search for interferences and
possible reasons if applicable.
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
safety, installation and commissioning
protective earth resistance
isolation resistance and
leakage current
The allen key must be pulled out of the screw
immediately after usage due to risk of injury!
created 04.05.2012
printed 04.05.2012
accomplished
accomplished
accomplished
accomplished
accomplished
page: 95 of 191