Setting of OP 41 to value 1 affects the safety of the subject and third parties.
A downward movement of the elevation after pressing STOP can result in a jumpy
reaction and in severe injuries and contusions.
h/p/cosmos strongly recommends adhering to the standard setting of this option. Stop
means stop for all movements and must not result in activity or movement on the part of
the running machine. It is only allowed to change option 41 after a written documented
instruction of the user regarding the danger warnings and after a written confirmation by
the customer/user.
OP46
Display and setting of elevation for
ladder ergometer
(only
available
ergometers, not for treadmills)
OP47
Reset of oil and service interval
OP48
Maximum speed during maximum
activation of the frequency inverter
(applicable for digital RS485 and/or
analogue control 0...10 volt of the
frequency inverter)
© 2012 h/p/cosmos sports & medical gmbh
$
adjustable from 0 to 180 degrees
for
ladder
After service including lubricating of the treadmill and all maintenance
work, the oil and service intervals can be reset with option OP47.
The maximum frequency inverter speed is achieved if the MCU
(within UserTerminal) sets the control voltage (analogue voltage for
frequency inverter) at 10 V. This means maximum frequency of the
inverter.
The maximum speed during maximum frequency of the inverter is set
with the input of the correct device type at option number 03 (OP03)
of the administrator options. Option 48 is necessary for manual speed
calibration or special speed transformation or drive motors (special
speeds) only.
When changing OP 48, always pay attention to the correct setting of
the OP 05 (maximum speed), OP 11 (speed measurement; has to be
switched on) and OP 34 (speed calibration value).
Examples for the different special speeds and the appropriate
maximum inverter frequencies are to be found in the list in chapter
10.1 entitled "List of running machines types"
Make sure that the correct device type is set at administrator OP03
N
Go to administrator OP 05 and set maximum possible speed.
N
Set the device in the manual mode to the maximum possible speed
N
by using key "+".
Check the analogue voltage at the MCU-FU control with a volt
N
meter. It should be 10 V. Make a note of the voltage in the service
report (not possible with digital control via RS485)!
Read the speed and make a note of the value in the h/p/cosmos
N
service report: OP 48: _____ km/h
Stop the running belt with the STOP key.
N
Go to administrator OP 48 and insert the previously measured
N
value in km/h (maximum possible speed at full FU-activation).
Go again to administrator OP 05 and indicate the desired /
N
appropriate maximum speed.
Info 1: If no speed scan (speed sensor) is set under OP11, the speed
calibration value will be calculated automatically with OP48.
Info 2: Even if no speed scan is set, the speed calibration value is of
internal significance. In that case the scanning impulses (actual
impulse) are deduced or simulated directly from the analogue voltage
and not measured via the speed sensor input.
Range of adjustment: 1.00 to 99.99 km/h
Fixed values selectable with up- /down-key: 20; 22; 22.5; 26; 30; 30.3;
40; 40.9; 41.2; 45.3; 46; 50; 60; 80
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
optional settings: administrator options
indicates elevation of ladder ergometer, standard: 75 degree,
created 04.05.2012
printed 04.05.2012
page: 128 of 191

Advertisement

Table of Contents
loading

Table of Contents