Page 1
The copyright of this document, media, design, software and intellectual property is owned by run ahead of time ® h/p/cosmos sports & medical gmbh, except where explicitly stated differently. Warning: All rights reserved. Unauthorized copying, reproduction, hiring, lending, public performance and broadcasting prohibited.
Page 3
These instructions for use include all of the information needed to operate the device correctly and safely. Please read them carefully before use and keep them available at all times. We hope you will have a lot of fun and success as you work with your h/p/cosmos device. Franz Harrer President & CEO h/p/cosmos sports &...
1.4 Labels on device In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels! Illustration...
The treadmill is purely used as a stressing device and training equipment. * Devices marked with an “r” or "rs" like “h/p/cosmos saturn 250/75 r” are intended for applications with wheels as well. Applications with wheels include cycling, roller skiing, wheelchair applications, etc...
❚ * Devices marked with an “r” or "rs" like “h/p/cosmos saturn 250/75 r” are intended for applications with wheels as well. Applications with wheels include cycling, roller skiing, wheelchair applications, etc... 3.2 Intended operator Persons carefully trained according to these instructions for use.
* Devices supplied with the special “running belt for ski and spike applications” (see “Annex III (Accessories)”) are intended for application with spikes or studs as well. ** Devices marked with an “r” or “rs” like “h/p/cosmos saturn 250/75 r” are intended for applications with wheels as well. Applications with wheels include cycling, roller skiing, wheelchair applications, etc.
Page 17
❚ mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels! In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service ❚ personnel via telephone and writing.
4 Safety (sport) h/p/cosmos medical treadmills may be operated with healthy subjects as well. The safety notes, warnings and precautions have to be pointed out to every user and operator and displayed within sight of the running machine. 4.1 Safety information – Forbidden use...
Page 19
❚ mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels! In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service ❚ personnel via telephone and writing.
However, the risk of injury or even death due to a malfunction of the treadmill is very low. In over 30 years of history and with more than 10,000 h/p/cosmos treadmills on the worldwide market, there has never been such a reported incident.
6.2 Standard vs. “lt” devices Most h/p/cosmos treadmills are available as standard or “lt” (light) devices. “lt” devices have no UserTerminal (no display, no keyboard). Standard device “lt” device Control via Control via - UserTerminal - Software (see “accessories”) - Software (see “accessories”) - additional keyboard (see “accessories”)
In case of any visible or assumed defects or malfunctions (of the device, accessories, software, etc.), unplug device, exclude ❚ reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing. Start and load changes within the automatic modes are indicated by acoustic signals (beeps).
Optional (*) At least one kind of fall prevention is mandatory, see “safety”. 9.3 Compatible devices A number of devices as well as software products are compatible with h/p/cosmos treadmills via coscom v3 interface protocol. Please contact service@hpcosmos.com for the list of compatible devices.
❚ mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels! In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service ❚ personnel via telephone and writing.
For further and precise analysis a detailed instructions regarding the tissue lubrication test are available at service@hpcosmos.com. Fill the syringe with 10 ml of the lubricant (h/p/cosmos special oil, only) and attach the tube to the syringe. Pump the lubricant under the running belt very slowly and remove the syringe.
Page 49
Start the running belt with 2 km/h and walk on the belt for 2 min. Vary the position in order to distribute the lubricant. Reset the OIL message (see OP01 in “User options”). Devices with automatic lubrication (devices marked with an “r” or “rs” like “h/p/cosmos saturn 250/75 r”) Description Illustration When the “oil tank”...
All maintenance work that is not explained in detail, must not be performed by the operator. Safety inspections, installation and repair work, must also not be performed by the operator. This kind of work must be performed by h/p/cosmos service personnel according to the “h/p/cosmos service instructions”. The “h/p/cosmos service instructions” are available at service@hpcosmos.com.
In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection, ❚ mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels! Problem...
12.2 RS232 troubleshooting Problem Solution No connection via RS232 For connection to PC with h/p/cosmos software and most external devices use the included RS 232 interface connection cable [cos00097010034]. (wrong cable) No connection via RS232 Check cable and plugs for defects.
Limited liabilities apply: If h/p/cosmos or h/p/cosmos organizational bodies, senior management or agents can be held accountable for the payment of damages pertaining to slight negligence (breach of material contractual obligations), the damages shall be limited to damages that could typically have been foreseen. Liability pertaining to slight negligence excludes liability as a result of loss of production, interruption of business and loss of profits.
17 Annex I 17.1 Installation This device must only be unpacked, transported, installed, commissioned, instructed, maintanined and repaired by h/p/cosmos service personnel (see maintenance). If the packaging has been damaged, please contact service@hpcosmos.com immediately. Make sure no objects, sand, stones, liquids, towels, jewellery, cell phones, containers with liquid etc. can fall into the device or onto the running surface or underneath the running belt.
17.2 Commissioning and instruction protocol When installing the device, the h/p/cosmos service personnel instructs the intended operator according to the instructions for use, following this instruction protocol. With the name and signature on the commissioning and instruction protocol, the customer confirm that the installation, commissioning and instruction has been performed correctly and instructed persons fully know how to operate the device safely.
Furthermore, this will result in loss of liability and warranty. Please fill out the instruction protocol and send it back to h/p/cosmos by fax (+49 18 05 16 76 69), email (sales@hpcosmos.com)
Page 65
“accessories / compatible devices”. OFF = RS232 not active / no protocol / interface deactivated h/p/cosmos coscom v1, v2, v3 with baud rate 9600 bps (standard setting COM 1 and COM 2) " [...] [...] " TM treadmill emulation in km/h...
Title Description Short description The h/p/cosmos para control is designed for remote control of medical devices. But it does not perform medical diagnosis or evaluation. The read out data shall not be the basis for diagnosis or evaluation. For remote control of all h/p/cosmos running machine and ladder-ergometers with MCU2, MCU3, MCU4 and MCU5.
Additional keyboard for h/p/cosmos treadmills [cos100815, cos100680] Title Description Short description Additional keyboard (no display!) with magnet holder for remote control of h/p/cosmos treadmill ergometer or h/p/cosmos ladder ergometer. Illustration Application The additional keyboard allows full control over all functions of the MCU like:...
Safety arch for treadmill family 200-300/75-125 [cos10171, cos10172, cos10173] Title Description Short description The h/p/cosmos safety arch is one option to secure the subject from falling. Furthermore, the safety arch stops the treadmill in case of falling. Illustration Application See “fall prevention”...
Deactivate brakes of the wheelchair (reactivate after treadmill use). Start application. Additional safety information Only devices marked with an “r” or “rs” like “h/p/cosmos saturn 250/75 r” are intended ❚ for applications with wheels as well. Applications with wheels include cycling, roller skiing, wheelchair applications, etc.
Deactivate brakes of the bicycle (reactivate after treadmill use). Start application. Additional safety information Only devices marked with an “r” or “rs” like “h/p/cosmos saturn 250/75 r” are intended ❚ for applications with wheels as well. Applications with wheels include cycling, roller skiing, wheelchair applications, etc.