h/p/cosmos 150/50 mercury Operating And Service Manual page 127

Running machine series
Table of Contents

Advertisement

OP37
Interval for service message TSC
(time-dependent)
for technical safety control
E.02 "HELP"
OP38
Service interval
(distance-dependent)
"E 02" "HELP"
OP39
Elevation: display as set point or
true value
(This option is only available for
treadmills, not for ladderergometers)
OP40
Speed: display as set point or true
value
OP41
Elevation control after STOP
(This option is only available for
treadmills, not for
ladderergometers).
© 2012 h/p/cosmos sports & medical gmbh
On sports devices and/or active medical devices with power supply
connection used for either commercial, industrial, institutional and
medical applications, regular technical safety controls have to be
performed by a trained and authorised service technician. Appropriate
protocols and inspection instructions are available at h/p/cosmos
service. An error message "E 02" "HELP" for the TSC-service interval
will appear in the display after expiration of a time period of 12
months.
The remaining time period (in months) till the next TSC / maintenance
/ inspection can be viewed by calling up OP37.
Reset of the interval and the error report „E02" is possible via option
OP47.
The TSC report interval can be changed via OP37:
Select OP37 and change with a or d. Confirm with c
OP37 can also be used for the adjustment of the maintenance
intervals in months if the mileage is small.
§
Adjustment range 0 to 48 months
h/p/cosmos recommends regular maintenance (interior cleaning,
checking drive belts, etc.) carried out by an authorized technician at
least every 5,000 km (for ladder ergometers: 500 km). For this
reason, the error message "E 02" "HELP" is displayed every 5000 km
(500 km). For detailed information, descriptions and specifications
about maintenance, please contact the h/p/cosmos service
department. An error message "E 02" "HELP" for the maintenance
interval will appear. TSC controls are to be performed at least every
12 months.
Reset of the interval and the error report „E02" is possible via option
OP47.
The service maintenance interval can be changed via option OP38.
Call up OP38 and change with
§
Adjustment range 1,000 to 9,900 km or 100 to 5,000 km for the ladder
ergometer.
0 = Display of elevation as set point
1 = Display of elevation as true value rsp. flashing set point report, if
the set point was changed with the keys and has not yet been
reached (report: current elevation angle; recommended default
setting).
0 = Display of speed as set point (report: target speed)
1 = Display of speed as true value rsp. flashing set point report if the
set point was changed with the keys and has not yet been reached.
(report: current speed; recommended default setting)
0 = Elevation remains at the present level after pressing the STOP
button (recommended; default-setting)
1 = Elevation decreases to 0 % after pressing the STOP button.
file: n:\article\cos14310m5-v1_06hpc-en\20120504_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.docx
email@h-p-cosmos.com
optional settings: administrator options
indicates: TSC-Interval in months (standard = 12 months)
a
reports: service interval in km (standard = 5.000 / 500)
created 04.05.2012
printed 04.05.2012
.
d
c
.
or
. Confirm with
page: 127 of 191