11 Maintenance
Obey the maintenance intervals claimed in chapter "maintenance".
❚
Obey the competences claimed in chapter "maintenance".
❚
A second person has to be present during maintenance.
❚
In case of any visible or assumed defects or malfunctions (of the device, accessories, software, etc.), unplug device, exclude
❚
reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing.
In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection,
❚
mark clearly and inform h/p/cosmos service personnel via telephone and writing. Do not change or remove any labels!
In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service
❚
personnel via telephone and writing.
Do not perform maintenance during use.
Proper maintenance is an important pre-condition for safety, reliability, function and accuracy of the device.
Support h/p/cosmos service personnel with the documents needed.
11.1 Intervals and competences
Maintenance
Daily inspection
Lubrication
Adjustment of running belt
Safety inspection
Change of safety arch rope
Tightening of the running belt
Adjustment of levelling sockets
Installation and repair work
h/p/cosmos recommends entering into a service contract with an authorized h/p/cosmos service technician.
A service contract provides the best preventive maintenance and care for the device.
The service contract is available at service@hpcosmos.com.
To receive information on becoming h/p/cosmos service personnel, please contact service@hpcosmos.com.
11.2 Daily inspection
Perform a daily visual inspection (see chapter "maintenance").
❚
Before daily use, check the whole device for wear and tear.
Description
Pay special attention to components with high probability of wear
and tear:
1. Running belt and non-slip surfaces
2. External cabling
3. All textile parts
4. Fall prevention incl. safety harness, buckle, carabiner, rope
and rope arrest
If there is any visible or assumed wear and tear, unplug device,
exclude reconnection and call h/p/cosmos service personnel.
© 1988 - 2019 h/p/cosmos sports & medical gmbh
Interval
Daily
When OIL message occurs
If due
12 months
24 months
If due
If due
If due
cos105000_170-190_65, rev. 1.03
Competence
Operator
Operator
Operator
h/p/cosmos service personnel only
h/p/cosmos service personnel only
h/p/cosmos service personnel only
h/p/cosmos service personnel only
h/p/cosmos service personnel only
Illustration