Power Connection

If the factory-supplied flexible power cord is not long enough or is not compli-
ant with local standards, a 3-meter European-type harmonized power cord is
also included.
Ensure that the area is safe and replace the power cord as follows:
1. Disconnect the power from the analyzer and remove the Terminal Cover
2. Unplug the Power Connector (Item 8 in Figure 2-7) and unscrew the two
3. Remove the ground wire from the threaded chassis stud (Figure 2-8).
Figure 2-7.
Remove 1 (L) and 2 (N).
4. Loosen and remove the power cord strain relief cap with compression
5. Prepare the new power cord using the original cord as a guide:
6. Feed the strain relief cap with compression bushing onto the new cord and
2-14 | 5830 Moisture Analyzer
from the Rear Panel.
(2) plug screws that secure the "1" (LINE) and "2" (NEUTRAL) wires in place.
10
LINE VOLTAGE
FUSE
TB1
POWER
CONNECTOR
L
N
G
8
1
2
3
1 (L) (LINE)
2 (N) (NEUTRAL)
bushing (Item 1 in Figure 2-2) and pull out the power cord.
a. Remove the outer jacket.
b. Measure the length of the wires.
c. Strip lengths of wires.
d. Ring connector termination, etc.
tighten securely as per original. If the new cord is a smaller diameter than
the original, consider using the alternate bushing that is included with the
European cord. Ensure that the power cord does not slip when pulled or
pushed from the outside.
TERMINATOR
TB2
ANALOG
OUTPUT
DATA
CONC
SYSTEM
RD
VALID
ALARM
ALARM
-
+
+
OUT
IN
1
2
3
4
5
6
7
8
9
10
11
PN 583054901, Rev YA
RS-485
ON
ON
TB3
TD
-
+
-
12
13
14

Advertisement

Table of Contents
loading

This manual is also suitable for:

583054901

Table of Contents