DVPlus Digital Viscometer
M21-400 REV.1.1
D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L

Advertisement

Table of Contents
loading

Summary of Contents for Ametek BROOKFIELD DVPlus

  • Page 1 DVPlus Digital Viscometer M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 2 PROPRIETARY RIGHTS NOTICE This manual contains valuable information and material developed by AMETEK Brookfield for use with the DVPlus Digital Viscometer. No part of this manual can be reproduced or transmitted in any form or by any means, electronic, mechanical or otherwise. This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of AMETEK Brookfield.
  • Page 3: Table Of Contents

    Table of Contents 1. INTRODUCTION ............................5 1.1 Components ................................6 1.2 Utilities ..................................7 1.3 Component Diagram ...............................9 1.4 Specifications ................................11 1.5 Installation .................................. 11 1.6 Safety Symbols and Precautions ........................12 1.7 Key Functions ................................13 1.8 Preventative Maintenance and Cleaning ....................... 14 2.
  • Page 4 APPENDIX D - SPINDLE ENTRY CODES AND RANGE COEFFICIENTS ......59 APPENDIX E - CALIBRATION PROCEDURES ................. 61 APPENDIX F - THE AMETEK BROOKFIELD GUARD LEG ............66 APPENDIX G - AUTOMATIC OSCILLATION CHECK ..............68 APPENDIX H - DVE 50-A PROBE CLIP ..................... 69 APPENDIX I - LABORATORY STANDS .....................
  • Page 5: Introduction

    The AMETEK Brookfield DV Viscometer series has provided exceptional value for viscometer users since its introduction in 1981. AMETEK Brookfield has continued to develop and improve to maintain its position in the market as the best value for QC applications. The DVPlus Viscometer continues in this tradition of innovation, quality, and value.
  • Page 6: Components

    Please check to be sure that you have received all components and that there is no damage. If you are missing any parts, please notify AMETEK Brookfield or your local authorized dealer immediately. Any shipping damage must be reported to the carrier. Please save packaging materials in case the instrument needs to be shipped for service.
  • Page 7: Utilities

    1.2 Utilities VAC; Hz Limits: 100 - 240 VAC; 50/60 Hz ±5% Power Supply: 24 Watts, Class II certified plug-in power supply Main supply voltage fluctuations are not to exceed 10% of the nominal supply voltage. Must be used with DPV-2019 Power Supply. Alternative power sources may cause damage to the instrument.
  • Page 8 KS C 9610-6-4:2017 Generic standards - Emission standard for industrial environments AS/NZS CISPR32 Electromagnetic compatibility of multimedia equipment - Emission requirements ETSI EN 301 489-1:2019-11 Ed. V2.2.3 Electromagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements; Harmonized Standard for Electromagnetic Compatibility ETSI EN 301 489-17:2017-03 Ed.
  • Page 9: Component Diagram

    1.3 Component Diagram DVPlus Viscometer Model G Laboratory Stand Shipping Cap Spindle Set Guard Leg Levelling Screw Temperature Probe Option • For use with SSA: • For use in beaker: Temperature Probe Clip DVE-50A DVE-50A Temperat Temperature Cable SC4-61Y Temperature Probe Temperature Figure 1-3.1 DVP-94Y...
  • Page 10 Figure 1-3.2 M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 11: Specifications

    Users within the European Union: Please contact your dealer or the local authorities in charge of waste management on how to dispose of this product properly. All AMETEK Brookfield offices and our network of representatives and dealers can be found on our website: www.brookfieldengineering.com.
  • Page 12: Safety Symbols And Precautions

    Assemble the Model G Laboratory Stand (refer to assembly instructions in Appendix I) 2. Attach the Viscometer head to the clamp on the laboratory stand rod. 3. Connect the RTD probe to the socket on the rear panel of the DVPlus. 4.
  • Page 13: Key Functions

    Refer to the manual for specific warning or caution information to avoid personal injury or damage to the instrument. Precautions: If this instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. This instrument is not intended for use in a potentially hazardous environment.
  • Page 14: Preventative Maintenance And Cleaning

    The Status Bar provides information relating to the date and time (as configured by the user) Status Bar and various connections to the DVPlus Viscometer. The Title Bar identifies the activity to be conducted in the current view and includes any Title Bar navigation icons that are relevant.
  • Page 15: Getting Started

    2. GETTING STARTED 2.1 Power up The DVPlus Viscometer will go through a Power Up sequence when the power is turned on. The Viscometer will present a black screen for approximately 4 seconds and then the Splash screen for 5 seconds. The Splash screen is shown below (Figure 2-1) and includes several critical parameters about the Viscometer including viscometer torque (LV, RV, HA, HB, or other), serial number, and the firmware version number.
  • Page 16: Autozero

    Call AMETEK Brookfield for additional assistance. TIP: When contacting AMETEK Brookfield or your local authorized dealer for technical support or repair services, please record the information on the Splash screen and the serial number (found on the serial tag located on the back of the instrument head) in any correspondence or shipping paperwork.
  • Page 17: Navigation

    to proceed. TIP: Do not touch the Viscometer during the AutoZero process to ensure the best zero value. TIP: The AutoZero sets the zero point of the viscometer range. A failure to properly level the Viscometer or to remove the spindle may affect the zero value and all measurement results. Figure 2-3.1 Figure 2-3.2 Figure 2-3.3...
  • Page 18: Spindle Selection

    Back Arrow Return to the previous screen. Home Icon Return to the Main Menu screen. Spindle Enter Spindle Selection screen. Speed Enter Speed Selection screen. End Condition Choose the desired end condition. QC Limits Choose the desired QC Limit Start Configured Test. Settings Various instrument settings.
  • Page 19 To remove the spindle, the lower shaft should be secured and slightly lifted with one hand while unscrewing the spindle to the right. Spindles can be identified by the number on the side of the spindle coupling nut. If your instrument has the magnetic coupling system, the spindles are attached as follows: Align two opposing spindle slots (Figure 2-5.2) with the two pins inside the coupling (Figure 2-5.3).
  • Page 20 Figure 2-5.6 Figure 2-5.7 The DVPlus requires a Spindle Entry Code number to calculate viscosity values. The two-digit code for each spindle can be found in Appendix D. Pressing the Spindle field on the Configure Viscosity Test screen will present the Set Spindle screen. Spindle Codes can be entered using a digital numeric keypad or a scrollable spindle list.
  • Page 21: Speed Selection

    Figure 2-5.9 If using the scrollable spindle list, press the dropdown button in the Spindle box to open the spindle list Scroll up or down through the list of available spindles until you find the spindle you want. Then simply touch the spindle name/part number to select it.
  • Page 22: Full-Scale Range

    Figure 2-6 If using the scrollable speed list, press the dropdown button in the Speed box to open the list of available pre-set speeds . Scroll up or down through the list until you find the pre-set speed you want, then simply touch it to select it.
  • Page 23: Display Resolution

    The DVPlus Viscometer offers a wide range of viscosity measurements. The data displayed will present a specific resolution depending upon the magnitude of the data. This resolution is consistent with the stated accuracy of the AMETEK Brookfield Viscometer. The table below shows the display resolution for viscosity, which will be utilized by the DVPlus Viscometer.
  • Page 24 • Beeper Volume Set the volume level for the internal beeper signal. • Screen Brightness Set the brightness level of the touch screen display. • Custom Spindles Enter the specifications for custom spindles. • Custom Speed Add a speed to the pre-set speed list. •...
  • Page 25 Temperature Offset Beeper Volume Screen Brightness Figure 2-9.4 Figure 2-9.5 Figure 2-9.6 Custom Spindles Custom Speeds Language Figure 2-9.7 Figure 2-9.8 Figure 2-9.9 M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 26 Reset Settings Set Date & Time Figure 2-9.10 Figure 2-9.11 Firmware Update Bluetooth Setup Figure 2-9.12 Figure 2-9.13 M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 27 NOTE: USB Flash Drives must be formatted to FAT32 for DVPlus Firmware Updating Insert a USB Flash Drive (FAT32 Format) with the Firmware Update into the USB A Port on the back of the DVPlus. DO NOT Remove the USB Flash Drive at any point during the upgrade. Temperature Probe USB B USB A...
  • Page 28 Select the desired Firmware version from the USB Drive. Figure 2-9.16 DVPlus will have you confirm your choice. Figure 2-9.17 M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 29: End Conditions

    Press Update and the DVPlus will automatically update the firmware to the desired selection. When the update is complete, the DVPlus will automatically reboot. DO NOT Remove the USB Flash Drive at any point during the update. Figure 2-9.18 Verify that the desired Firmware version has been installed by viewing the startup Splash Screen. Figure 2-9.19 After the DVPlus finishes booting, the USB Flash Drive can be removed.
  • Page 30: Qc Limits

    A short time value may result in erroneous data if the torque sensor has not achieved a stable deflection for the measurement condition. AMETEK Brookfield recommends a minimum time value of 20 seconds. However, this can be reduced based on user judgement by observing when the torque stabilizes.
  • Page 31: Viscosity Wizard

    To set a QC Limit, press the QC Limits box on the Configure Viscosity Test screen. The Set QC Limits screen will then be displayed. (Figures 2-11.1 & 2-11.2). Use the keypad to enter the desired low and high limits. The minimum Low Value for all parameters is 0 except Temperature, which can be set as low as -100°C.
  • Page 32 Example Liquid/Estimated Value is outside of your DVPlus optimal torque range, or if Non-Flowing is selected, the DVPlus will prompt the user to contact Ametek Brookfield to discuss options for measuring your fluid or non-flowing materials (see Figure 2-12.4).
  • Page 33 Choose Spindle: There are two options for spindle selection. The first option is that the Wizard will recommend a spindle to be used based on the torque range of your DVPlus and the Estimated Viscosity of your fluid. The second option will allow a spindle to be selected by manually entering the spindle code (see Figure 2-12.5). 2-12.5 Once the desired spindle is selected, press the Next key to move on to the “Determine Speed”...
  • Page 34 2-12.9 If the Viscosity Wizard has not been able to obtain torque measurements within 20-80% after three speed changes, the “Please Contact AMETEK Brookfield” screen will be displayed (see Figure 2-12.10). M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 35 2-12.10 If the Viscosity Wizard obtains a torque measurement between 20-80%, the DVPlus may adjust the speed in order to get as close to 50% torque as possible. When the best torque is reached, the DVPlus will display the “Wizard: Results”...
  • Page 36: Custom Spindle

    A Custom Spindle can be created when the user has developed a new spindle geometry or when using a standard AMETEK Brookfield spindle in a non-standard measurement container. The Custom Spindle, once created, will be available to users through the Spindle selection in Configure Viscosity Test.
  • Page 37 M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 38: Making Viscosity Measurements

    3. MAKING VISCOSITY MEASUREMENTS 3.1 Quick Start 3.1.1 Instrument Setup • Assemble the Model G Laboratory Stand (refer to assembly instructions in Appendix I of this manual. • Insert the instrument mounting rod into the lab stand clamp and tighten. •...
  • Page 39: Preparations For Making Measurements

    Brookfield with specific sample chambers such as the Small Sample Adapter, UL Adapter, and Thermosel. AMETEK Brookfield recommends that you use the appropriate container for the selected spindle. You may choose to use an alternate container for convenience. However, this may have an effect on the measured viscosity. The DVPlus is calibrated considering the specified container.
  • Page 40: Selecting A Spindle/Speed

    Figure 3-2.1 Use the Temperature Probe Clip to secure the Temperature Probe to the rim of your Sample Beaker/Container or Guard Leg Figure 3-2.2 3.3 Selecting a Spindle/Speed The DVPlus has the capability of measuring viscosity over a wide range. For example, the DVPlus RV-range can measure fluids within the range of 100-40,000,000 cP (see Appendix A).
  • Page 41: Running A Test

    speed, container, temperature, and test time. 3.4 Running a Test A viscosity test is started by pressing the Run button while viewing the Configure Viscosity Test screen. Live measurement data will be displayed during the test. When the end condition is met, the final data will be displayed on the DVPlus until the user presses a key.
  • Page 42: Communication With Optional Dvplus Connect

    3.5 Communication with Optional DVPlus Connect The DVPlus Viscometer can be used in conjunction with the AMETEK Brookfield software program DVPlus Connect App. DVPlus Connect App will collect the data output from the DVPlus and allow for data storage and data printing.
  • Page 43: Connecting To The Dvplus Connect App

    Instrument Connection Page Instrument Connection Page allows a user to connect to an instrument via Bluetooth® or USB-B cable. When loaded, the page scans for and displays available Bluetooth connections. The User has the option to disconnect, rescan, or connect to a previously connected instrument.
  • Page 44: Establishing Bluetooth® Connection To The App For A First Time

    3.5.2 Establishing Bluetooth® connection to the app for a first time • Make sure that the PC and DVPlus instrument are next to each other. • On your PC, make sure that Bluetooth® connection is enabled: a. Windows 11 • On the taskbar, select the Network icon.
  • Page 45 Figure- 6 • Here you can see screenshots of a successful connection. Figure- 7 & 8 • Upon successful connection, the DVPlus Connect App will switch to the Live Data Page. The connected DVPlus’ torque range and serial number will appear at the top of the app (see Figure 9). M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 46: Establishing Bluetooth® On Regular Basis

    Figure- 9 • Once pairing is established, the status of the DVPlus under Bluetooth® Connection will change to Paired: True (see Figure 10). 3.5.3 Establishing Bluetooth connection to a Paired Instrument • Make sure the DVPlus instrument is in the visible range, or no more than 50 ft. away. •...
  • Page 47: Establishing Bluetooth® On Regular Basis

    3.5.4 Setting up Bluetooth Automatic reconnection DVPlus Connect App can be set up to reconnect with a DVPlus instrument if: • Instrument was out of the range after Bluetooth connection was established. • Instrument was restarted after Bluetooth connection was established. To enable this feature, navigate to the Instrument Connection page and check the Attempt to Reconnect if connection is lost option.
  • Page 48: Appendix A - Viscosity Ranges

    APPENDIX A - VISCOSITY RANGES VISCOSITY RANGE TABLES Viscosity ranges shown are for operational speeds 0.1 through 200 RPM. LV Viscometer with LV spindles #1-4 and RV/HA/HB Viscometers with spindles #1-7 Viscosity Range cP Viscometer Minimum Maximum DVPlusLV 6,000,000 DVPlusRV 40,000,000 DVPlusHA 80,000,000...
  • Page 49 DIN Adapter Accessory DAA Spindle Viscosity (cP) Shear Rate Sec-1 DVPlusLV DVPlusRV DVPlusHA DVPlusHB 0.6 - 5,000 6.1 - 5,000 12.2 - 5,000 48.8 - 5,000 1.29N 1.8 - 10,000 18.2 - 10,000 36.5 - 10,000 146 - 10,000 1.29N 5.7 - 50,000 61 - 50,000 121 - 50,000...
  • Page 50 Vane Spindles Spindle Torque Range Shear Stress Range Viscosity Range (Pa) cP(mPa•s) V-71 NOT RECOMMENDED FOR USE ON LV TORQUE V-72 0.188-1.88 104.04-1.04K V-73 0.938-9.38 502-5.02K V-74 9.38-93.8 5.09K-50.9K V-75 3.75-37.5 1.996K-19.96K V-71 0.5-5 262-2.62K V-72 2-20 1.11K-11.1K V-73 10-100 5.35K-53.5K V-74 100-1K...
  • Page 51 (including the critical parameters of temperature control and stated fluid accuracy). AMETEK Brookfield’s accuracy statement of +/- 1% of the range in use applies to the AMETEK Brookfield rotational Viscometer...
  • Page 52: Appendix B - Variables In Viscosity Measurements

    RPM to a high RPM and then back to the low RPM, and the readings are lower for the descending step, the material is time dependent, thixotropic. AMETEK Brookfield publication, “More Solutions to Sticky Problems”, includes a more detailed discussion of rheological properties and non-Newtonian behavior.
  • Page 53 4. Viscometer model 5. Spindle used 6. Whether or not to attach the guard leg Test speed or speeds (or the shear rate) 8. Length of time or number of spindle revolutions to record viscosity 9. How the sample was prepared and/or loaded into the container M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 54: Appendix C - Spindle Entry Codes And Smc/Src Values

    APPENDIX C - SPINDLE ENTRY CODES AND SMC/SRC VALUES Each spindle has a two-digit entry code which is entered via the keypad on the DVPlus. The entry code allows the DVPlus to calculate Viscosity, Shear Rate, and Shear Stress values. Each spindle has two constants that are used in these calculations.
  • Page 55 SPINDLE ENTRY CODE 0.64 1.223 DIN81 1.29 DIN82 3.75 1.29 DIN83 12.09 1.29 DIN85 1.22 1.29 DIN86 3.65 1.29 DIN87 12.13 1.29 SC4-14 SC4-15 0.48 SC4-16 0.29 SC4-18 1.32 SC4-21 0.93 SC4-25 0.22 SC4-27 0.34 SC4-28 0.28 SC4-29 0.25 SC4-31 0.34 SC4-34 0.28...
  • Page 56 Table D-2 lists the model codes and spring torque constants for each Viscometer model. Table D-2 Model Model Code on DVPlus Screen DVPlusLV 0.09375 DVPlusRV DVPlusHA DVPlusHB The full-scale viscosity range for any DVPlus model and spindle may be calculated using the equation: Full-Scale Viscosity Range Where: = DVPlus Torque Constant from Table D-2...
  • Page 57 Table D-4 PRE-SHEAR SPEEDS (RPM) Range Increment 0.1 to 59.9 60 to 200 ZERO STEP SPEEDS (RPM) RUN SPEEDS (RPM) RANGE INCREMENT RANGE INCREMENT 0.1 to 59.9 0.1 to 59.9 The shear stress measurement range for the three standard (or supplied) vane spindles at each spring torque is as follows: Table D-5 SPINDLE...
  • Page 58 Where: θ = Radial spring factor (rad/%torque) = Spring windup angle (revolutions) Where: θ = Spring windup angle (revolutions) = Calibration speed (fixed at 0.1 rpm) (rpm) = Calibration time (time for 0% to 100% spring windup) (milliseconds) Where: = Calibration time (time for 0% to 100% spring wind up during calibration) (milliseconds) = Base increment (milliseconds)
  • Page 59: Appendix D - Spindle Entry Codes And Range Coefficients

    APPENDIX D - SPINDLE ENTRY CODES AND RANGE COEFFICIENTS The range coefficient is a convenient tool for quickly determining the maximum viscosity that can be measured with a specific spindle/speed combination. Identify the spindle in use and the torque range (LV, RV, HA, HB) of the Viscometer.
  • Page 60 Range Coefficient Spindle Entry Code 187,200 2,000,000 4,000,000 16,000,000 468,000 5,000,000 10,000,000 40,000,000 936,000 10,000,000 20,000,000 80,000,000 Spiral 98,400 1,050,000 2,100,000 8,400,000 6,400 12,800 51,200 HT-DIN-81 3,420 36,500 73,000 292,000 SC4-DIN-82 3,420 36,500 73,000 292,000 SC4-DIN-83 11,340 121,300 242,600 970,400 ULA-DIN-85 1,144 12,200...
  • Page 61: Appendix E - Calibration Procedures

    APPENDIX E - CALIBRATION PROCEDURES The accuracy of the DVPlus is verified using viscosity standard fluids which are available from AMETEK Brookfield or your local authorized dealer. Viscosity standards are Newtonian, and therefore, have the same viscosity regardless of spindle speed (or shear rate). Viscosity standards, calibrated at 25°C, are shown in Table F-1 (Silicone Oils) and Table F-2 (Mineral Oils).
  • Page 62 AMETEK Brookfield does not re-certify Viscosity Standard Fluids. We will issue duplicate copies of the Certificate of Calibration for any fluid within two years of the purchase date. AMETEK Brookfield Viscosity Standard Fluids are reusable provided they are not contaminated. Normal practice for usage in a 600 mL beaker is to return the material from the beaker back into the bottle.
  • Page 63 CALIBRATION PROCEDURE FOR A THERMOSEL SYSTEM AMETEK Brookfield recommends a two-step check. First verify the calibration of the Viscometer using the standard Viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6) as detailed in this Appendix. Second verify the calibration of the Viscometer using the Thermosel.
  • Page 64 Calibration Test Results). However, instrument accuracy is ±2% of the maximum viscosity range and not the standard 1%. Viscometers as shown in Table E-2. Consult with AMETEK Brookfield or an authorized dealer to determine which fluid is appropriate. INTERPRETATION OF CALIBRATION TEST RESULTS When verifying the calibration of the DVPlus, the instrument and viscosity standard fluid error must be combined to calculate the total allowable error.
  • Page 65 Example: Calculate the acceptable accuracy for viscosity measurement using DVPlus LV with SC4-21 spindle in Small Sample Adapter at 6, 12, and 30 RPM. AMETEK Brookfield viscosity standard fluid 100 cPs has an actual value of 101.5 cP at 25ºC.
  • Page 66: Appendix F - The Ametek Brookfield Guard Leg

    AMETEK Brookfield. The guard leg is a part of the calibration check of the AMETEK Brookfield LV and RV series Viscometer. Our customers should be aware of its existence, its purpose, and the effect that it may have on data. With this knowledge, the Viscometer user may make modifications to the recommended method of operation to suit their needs.
  • Page 67 B-21Y B-20Y Guard Leg For RV Torque Guard Leg for LV Torque Figure G-1 M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 68: Appendix G - Automatic Oscillation Check

    Select the Oscillation Check from the Device Setup menu and follow the instructions on the touch screen display. When the Oscillation Check is complete, the touch screen display will report results as either Pass or Fail. If the Check passes, then press OK to continue. If the Check fails, contact your authorized AMETEK Brookfield representative for assistance.
  • Page 69: Appendix H - Dve 50-A Probe Clip

    APPENDIX H - DVE 50-A PROBE CLIP Probe Clip DVE-50A is supplied with all model DV series Viscometers, Rheometers, and Digital Temperature Indicators. It is used to attach the RTD temperature probe to the LV Guard Leg (Part No. B-20Y) or 600 mL low form Griffin beaker.
  • Page 70: Appendix I - Laboratory Stands

    APPENDIX I - LABORATORY STANDS Model G is the standard laboratory stand that comes with the DVPlus Viscometer. Figure K-1: Model G Laboratory Stand Item Part No. Description VS-CRA-14S Upright Rod and Clamp Assembly VS-CRA-18S Upright Rod and Clamp Assembly Optional GV-1201 Base, includes 2 GV-1203 leveling screws...
  • Page 71 Model QB is an optional laboratory stand which can be ordered for use with the DVPlus Viscometer. The advantage is the rapid speed of movement for lowering and raising the Viscometer head. Figure K-2: Model QB Laboratory Stand Item Part No. Description VSQA-100Y Upright Rod and Clamp Assembly...
  • Page 72 Do not use the DVPlus Viscometer with any laboratory stand that does not utilize the GV-1201 base. This large base is necessary for stability of the DVPlus Viscometer during use. Earlier versions of the AMETEK Brookfield Laboratory Stand including the Model A and Model S should not be used with the DVPlus.
  • Page 73: Appendix J - Screen Protector

    For additional protection of the DVPlus device, we recommend the use of a Touch Screen Protector Kit. The 5" screen protector prevents scratches and smudges. For more information on the AMETEK Brookfield Touch Screen Protector Kit, refer to the image below. M21-400 REV.1.1...
  • Page 74: Appendix K - Fault Diagnosis And Troubleshooting

    APPENDIX K - FAULT DIAGNOSIS AND TROUBLESHOOTING Listed are some of the more common problems that you may encounter while using your Viscometer. SPINDLE DOES NOT ROTATE • Make sure the Viscometer is plugged in. • Check the voltage rating on your Viscometer (100-240 V); it must match the wall voltage. •...
  • Page 75 Check for erratic spindle rotation. • Verify power supply. • Contact AMETEK Brookfield or your local authorized dealer for repair (see Appendix M). • Bent spindle or spindle coupling. • Contact AMETEK Brookfield or your local authorized dealer for repair (see Appendix M).
  • Page 76: Appendix L - Online Help And Additional Resources

    AMETEK Brookfield has its own YouTube channel. Videos posted to our website can be found here as well as other "homemade" videos made by our own technical sales group.
  • Page 77: Appendix M - Warranty Repair And Service

    All AMETEK Brookfield DVPlus Viscometers are supplied from the factory with a Calibration Seal ( Located on the back of the viscometer). The warranty stated above will be voided if the Calibration Seal has been damaged. Only AMETEK Brookfield or our authorized servicing dealers may break the Calibration Seal for purposes of instrument repair or recalibration.
  • Page 78 For repair or service in China, return to: AMETEK Commercial Enterprise (Shanghai) Co. Ltd. Room 810 Dongbao Plaza, No.767 East Dongfeng Road, Guangzhou, China (510600) Tel: 86-20-8363 4768 Fax: 86-20-8363 3701 Email (Sales): brookfield-china.sales@ametek.com Email (Service): brookfield-china.service@ametek.com www.brookfieldengineering.cn For repair or service in Germany, return to: AMETEK GmbH / B.U.
  • Page 79 Email (Sales): brookfield-thailand.sales@ametek.com Email (Service): brookfield-thailand.service@ametek.com store.brookfieldengineering.asia Onsite service at your facility is available from AMETEK Brookfield. Please contact our Service Department in the United States, United Kingdom, Germany, India, Thailand, or China for details. M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...
  • Page 80 This tear-off sheet is a typical example of recorded test data. Please photocopy and retain this template so that additional copies may be made as needed. M21-400 REV.1.1 D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L...